The Philips Air Hoses for NIBP Cuffs are designed for non-invasive blood pressure (NIBP) measurement in neonatal, pediatric, and adult patients. They are intended for use by, or under the supervision of, a licensed physician or other healthcare provider.
Function Description:
These air hoses connect NIBP cuffs to Philips NIBP monitoring devices, facilitating the transmission of air pressure for blood pressure measurement. The manual illustrates different types of air hoses, categorized by patient population (Neonatal, Adult, Mobile CL Adult Bayonet, and Efficia Adult) and length (1.0m, 1.5m, and 3.0m). Each hose type has specific reference numbers (REF) for identification.
Important Technical Specifications:
The air hoses are available in various lengths to accommodate different clinical setups:
- Neonatal NIBP Air Hose:
- M1596C: 1.5m Hose
- M1597C: 3.0m Hose
- Adult NIBP Air Hose:
- M1598B: 1.5m Hose
- M1599B: 3.0m Hose
- Mobile CL Adult Bayonet Air Hose:
- Efficia Adult NIBP Air Hose:
- 989803160881: 1.5m Hose
- 989803160891: 3.0m Hose
The hoses are designed for specific cuff sizes:
- Neonatal and Infant #5 cuffs are approved for use only with M1596C and M1597C hoses.
- Infant and Pediatric cuffs are approved for use only with M1598B, M1599B, Mobile CL, and Efficia hoses.
- Small Adult, Adult, Large Adult, and Thigh cuffs are approved for use only with M1598B, M1599B, Mobile CL, and Efficia hoses.
The product packaging includes symbols indicating:
- Quantity per package.
- "Follow operating instructions" (represented by an 'i' in a box).
- "Not manufactured with natural rubber latex" (a crossed-out latex glove symbol).
- "Not manufactured with polyvinyl chloride (PVC)" (a crossed-out PVC symbol).
- "Not manufactured with Di(2-ethylhexyl)phthalate (DEHP)" (a crossed-out DEHP symbol).
- "Caution" (an exclamation mark in a triangle).
- Storage temperature range (e.g., °C/°F with a temperature range).
- Storage humidity range (e.g., %RH with a humidity range).
- "Prescription use only" (Rx only symbol).
Usage Features:
The manual provides visual instructions for connecting and disconnecting the air hoses.
- For M1596C, M1597C hoses: The connection involves aligning and pushing the connectors together, then twisting to secure. Disconnection is achieved by twisting in the opposite direction and pulling apart.
- For M1598B, M1599B, Mobile CL, Efficia hoses: Connection involves aligning and pushing the connectors together until they click. Disconnection is achieved by pulling the outer sleeve of the connector back and then pulling the connectors apart. A "do not pull on the hose" symbol emphasizes proper disconnection technique.
- For Mobile CL hoses (specifically connecting to the monitor): The connection involves aligning the hose connector with the port on the monitor and pushing it in until it clicks. Disconnection involves pulling the outer sleeve of the connector back and pulling it out of the monitor port.
Maintenance Features:
The air hoses are not designed to be sterilized and should not be autoclaved. It is crucial to prevent cleaning or lubricating fluids from entering the hose, as this can damage the monitoring equipment.
For M1598B, M1599B, Mobile CL, and Efficia hoses, if connecting to cuffs becomes difficult, the cuff-end connector can be lubricated with a non-silicone spray release agent. However, instrument-end connectors and connectors for M1596C and M1597C hoses should not be lubricated.
The manual provides a comprehensive cleaning and disinfection table for all hose types, indicating approved cleaners/disinfectants:
- Approved for all hose types (Neonatal, Adult, Mobile CL, Efficia): Mild soap, Isopropyl alcohol, Hydrogen peroxide, Sodium hypochlorite, Viraguard®, TechSpray®, Sani-Cloth® Bleach, Sani-Cloth® HB, Super Sani-Cloth®, Carpe Diem™/MC Tb Wipes.
- Approved for Adult, Mobile CL, Efficia (but not Neonatal): Metrex Cavi-Wipes, Oxivir® Tb Cleaner Disinfectant, Oxivir® Tb Wipes, Bacillol AF, Meliseptol.
- Approved for Mobile CL and Efficia (but not Neonatal or Adult): Sani-Cloth® PLUS.
After cleaning and disinfection, the hoses should be wiped with clean water to remove all residue.
Warnings:
- Neonatal and infant #5 cuffs must only be used with hoses designated for Neonatal use.
- Pediatric and adult cuffs must only be used with hoses designated for Adult use.
- Philips NIBP air hoses use connectors that have the potential to misconnect with small-bore connectors used in non-NIBP healthcare applications. Users must ensure that connectors are properly connected to Philips NIBP devices only.