EasyManua.ls Logo

Philips PM 3266 - Page 152

Philips PM 3266
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
157
A
(B)
AMPL
Ya
(Yb)
input
signal
Adjuster
Trace
height
2 mV
12 mV C101
(ch. A)
or Cl
51 (ch. B)
3
DIV
t-/-3%
5
mV
30
mV
-
3DIV+/-3%
10 mV
60
mV
-
3DIV-t-/-3%
20 mV
120 mV
-
3DIV-h/-3%
50 mV
300
mV
-
3DIV-t/-3%
100 mV
600
mV Cl
04
(ch. A)
or Cl
54 (ch.
B)
3 DIV
-^/-3%
.2
V
1,2 V
3
DIV
+/-3%
.5 V
3 V
3
DIV +/-3%
1
V
6 V
Cl
09
(ch.
A) or
Cl
59 (ch. B)
3 DIV
-i-/-3%
2 V
12
V
-
3DIV+/-3%
5 V
30
V
3DIV-h/-3%
Remove the
input signal.
3.4.
10.4. Square-wave
response final Y
amplifier
Depress push-button
ALT of the display-mode controls (SI
)
Depress
push-buttons A and
B (COMP)
of the m.t.b. trigger-source controls
(S22)
Depress push-button MTB of the
X
deflection controls
(S2)
Set the
Y
POSITION controls
to
obtain a distance of 6 DIV
between both time-base
lines
(channel
A
time-base line
at
the
top)
Set the
X Masnifier
in the
x1
position
Remove the main time-base connector from the alternate
control pulse socket on the intermediate
amplifier board,
unit
12,
socket
9,
Fig. 3.24.
Connect
a
square-wave generator to
socket
9,
Fig. 3.24.
Set
the
output voltage of the
generator
to 3 V
Both time-base lines will be displayed at a
frequency
determined
by
the frequency of the square-wave
voltage.
200 Hz (unit
13)
Set
the generator frequency to 200
Hz
Set
the
m.t.b.
controls
to
obtain a
suitable,
triggered display
Check
that the
top
of
the displayed
pulse in straight within
2%;
if necessary,
put
R634
in its
mid
position and
select C613 to such
a
value that
the square wave response
is optimal.
Adjust R634
to optimal
square wave response.
2 kHz (unit
13)
Set the generator
frequency
to 2
kHz
Set the m.t.b.
controls to
obtain a
suitable, triggered display
Check that
the top
of
the displayed
pulse is straight within
2%;
if necessary,
put
R636
in its mid
position
and select C614 to such
a
value that the
square wave response
is
optimal.
Adjust
R636
to
optimal
square wave response.
Reconnect socket
9
(Fig.
3.24.).
3.4. 10.5.
Square-wave
response
channel
A
(unit 12)
Depress
push-button A of
the
display-mode
controls (SI)
Push the Y POSITION
controls to the NORM position
Depress
push-button DC of the channel A signal-coupling controls
Set
the channel
A
AMPL
switch
to
10 mV/DIV and its vernier to CAL
-
Set the XMagnifier in the XI position
Depress push-button
MTB
of the
Xdeflection controls (S2)
Depress push-button A of the
m.t.b.
trigger-source controls
(S22)
Set the m.t.b.
TIME/DIV switch
to a
suitable
value
Set the d.t.b. TIME/DIV switch to
OFF
Apply a
square-wave voltage of
60
mV, rise time 1 ns, repetition
rate 2
kHz,
to
the channel A input
Determine
(for example with the use
of an auxiliary
potentiometer)
the value for R256
that gives the
best square
wave response. A resistor
with
the
selected
value (between
1k6
and
4k7) must be
soldered
on the unit.
Check
that the pulse top
is straight
within 2% and that
the rise time is as
short as
possible; if
necessary
readjust
potentiometer R255A
or
select a
different value
for capacitor
C224 on the
intermediate
Y
amplifier adjusting
board
Increase the
repetition rate of
the input
signal
to
30
kHz
-
Put
the
instrument in the
Magnifier X10
mode for
better
waveform
analysis

Related product manuals