38
Fast electric
transients (Burst)
EN 61000-4-4:2012 (IEC 61000-4-4:2012)
(Test Levels: ±2 kV AC ports, ±1 kV Signal
ports)
Surges EN 61000-4-5:2014 (IEC 61000-4-5:2014)
(Test Levels: ±1 kV L-N)
Radio frequency
common mode
EN 61000-4-6:2014 (IEC 61000-4-6:2013)
(Frequencies: 0.15-80 MHz; Test levels: 10 V,
Modulation 1kHz, 80%AM)
Magnetic elds
(industrial
frequencies)
EN 61000-4-8:2010 (IEC 61000-4-8:2009)
(Frequencies: 50 Hz or 60 Hz, Test levels: 30
A/m)
Voltage dips and
interruptions
EN 61000-4-11:2004 (IEC 61000-4-11:2004)
(Test Levels: 0% at 0, 45, 90, 135, 180, 225,
270, 315° for 10,20,5000 ms /70% at 0°for500
ms; 0% at 0, 45, 90, 135, 180, 225, 270, 315°
for 0.5, 1, 250/300 cycles /70% at 0° for 25/30
cycles)
Variations of power
frequency
IEC 60601-1 (§ 4.10.2):2005
(Test levels: 50 +/- 1 Hz and 60 +/- 1 Hz)