6
DEFECT TYPE CAUSE SOLUTION
Scarce nebulisation Nebuliser cup clogged Remove all components of the
nebuliser cup, disconnect and
clean them complying with the
instructions provided in this in-
struction manual.
Scarce nebulisation Nebuliser cup clogged If the problem persists after clean-
ing, replace the nebuliser cup.
Nebulisation absent Nebuliser cup components not
assembled correctly
Take apart and reassemble the nebu-
liser cup; If the problem persists, re-
place the nebuliser cup.
Nebulisation absent The compressor does not work Be sure that the plug for the power
cord is disconnected from the
socket, check that the on/off switch
is in the "on" position and that the
electrical socket works.
Slow nebulisation Drug too thick Dilute the medicine as indicated
by the pharmaceutical company
TECHNICAL DATA
Nominal voltage: 230Vac
Service life in storage conditions: 5 years
Absorbed power: 140 VA
Frequency: 50 Hz
Maximum ow: 18 ± 4 l/min
Fuse: T1.6A L, 250 V, 5x20 mm
Max pressure: 2.2 ± 0.6 bar
Maximum pressure [internal tubing]: 3 bar
Operational ow: 7 ± 2 l/min
Conditions for using the device: 20 min ON - 40 min OFF at 40°C
Operational pressure: 0.70 ± 0.3 bar
NEB RATE: > 0.34 ml/min
MMAD: 3.08 µm ± 0.16 µm
Weight: 1.6Kg
Nebuliser cup capacity: 10ml Max; 2ml Min
Breathable fraction: 71.8% ± 2.21
Appliance service life: 1000 cycles/treatments
Noise level at 50 cm*: 63dBA*
Service life of accessories: 1 year
* Data on noise levels were measured on a new device. Values may vary over time with use.
The information on the unit’s capabilities provided by the manufacturer in accordance with EN 13544-1 may not
apply to medicines provided in suspension form or those that are highly viscous.
EN 60601-1 Medical electrical equipment – Part 1: general requirements for safety
EN 60601-1-2 Medical electrical equipment – Part 1: General requirements for basic safety and essential perfor-
mance - Collateral standard: electromagnetic compatibility - Requirements and tests.
EN 13544-1 Respiratory therapy equipment – Part one: Nebulising systems and their components
This section contains information specific to product compliance with the EN 60601-1-2 standard. PIC aerosol is a
medical electrical device that requires special precautions regarding electromagnetic compatibility and needs to
be installed and commissioned according to the electromagnetic information provided.