Mobile and portable RF communications equipment (mobile phones, transceivers, etc.) may affect the device.
AirFamily Evolution 02038208000000 is designed for use in an electromagnetic environment that
meets the specifications provided below. The person who purchased or who uses AirFamily Evolution
02038208000000 is responsible for ensuring that the device is used in an environment that complies with
said specifications.
Manufacturer’s Declaration and Guidelines – Electromagnetic Emissions
Phenomenon
Professional
healthcare
environments
a
)
HOME
ENVIRONMENT
a
)
Radiated and conducted radiofrequency EMISSIONS
a
)
CISPR 11
Group 1 Class B
Harmonic distortion IEC 61000-3-2
b
) Class A
Voltage fluctuations/flicker COMPLIANT WITH IEC 61000-3-3
b
)
a
) The device is intended for use in a home environment or professional healthcare structure and may
exclusively be used in hospital rooms and in sections of hospitals or clinics set up for respiratory therapy.
The strictest acceptance limits required for Group 1 Class B (CISPR 11) have been considered and applied.
The device is suited for use in the environments mentioned above so long as it is connected to the public
electric mains.
b
) The test may be applied to this environment so long as the ME EQUIPMENT and ME SYSTEM used are
connected to the PUBLIC ELECTRIC MAINS and the power supply is compliant with the standard applicable
policy regarding electromagnetic compatibility (EMC).
Manufacturer's Declarations and Guidelines - Electromagnetic immunity - Enclosure Door
Phenomenon
Applicable EMC policy
or test method
Immunity test levels
Professional
healthcare
environments
Home environment
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
± 8 kV in contact
± 2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated radio-frequency
electromagnetic fields
IEC 61000-4-3
a
)
10 V/m
b
) 80 MHz
- 2.7 GHz 80% AM
at 1 kHz
Fields near radio-
frequency devices for
wireless communication
IEC 61000-4-3 COMPLIANT
NB: further information about the distances to be
maintained from portable and mobile radio-fre-
quency communication devices (transmitters) and
AirFamily Evolution 02038208000000 may be