18
19
Phenomenon
Applicable EMC policy
or test method
Immunity test levels
Professional
healthcare
environments
Home environment
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
± 8 kV in contact
± 2 kV, ±4 kV ±, ±8 kV, ±15 kV air
Radiated radio-frequency
electromagnetic fields
IEC 61000-4-3
a
)
10 V/m
b
)
80 MHz - 2.7 GHz
80% AM at 1 kHz
Fields near radio-
frequency devices for
wireless communication
IEC 61000-4-3 COMPLIANT
NB: further information about the distances to be
maintained from portable and mobile radio-frequen-
cy communication devices (transmitters) and AirPro-
jet plus 02038204100000 may be requested by
contacting PIKDARE S.p.A. using the contact informa-
tion included in this booklet. It is nevertheless advis-
able to keep the electromechanical aerosol therapy
device at a safe distance (at least 0.5 m) from mobile
phones or other communication devices that use ra-
dio frequencies, to minimise interference.
Electromagnetic fields for
RATED mains frequency.
IEC 61000-4-8 30 A/m
c
)
50 Hz or 60 Hz
a
) The device is intended for use in a home environment or professional healthcare structure and may
exclusively be used in hospital rooms and in sections of hospitals or clinics set up for respiratory therapy. The
strictest limits for acceptance were considered and applied during immunity testing.
b
) Before modulation.
c
) This test level presupposes a minimum distance of at least 15 cm between the AirProjet plus
02038204100000 and the sources of magnetic fields at the mains frequency.
Manufacturer's Declarations and Guidelines - Electromagnetic immunity - Door for access with
alternating current
Phenomenon
Applicable EMC policy
or test method
Immunity test levels
Professional
healthcare
environments
Home environment
Fast electrical transient/
burst
IEC 61000-4-4 ± 2 kV
Repetition frequency at 100 kHz
Line to line voltage
impulses
IEC 61000-4-5 ± 0.5 kV, ±1 kV
Line to ground voltage
impulses
IEC 61000-4-5 Class II Device or equipped with Class II AC/DC
adapter. Test not applicable.
± 0.5 kV, ±1kV; ±2kV
Disturbances conducted
induced by radio-
frequency fields
IEC 61000-4-6 3 V
a
) 0.15 MHz - 80 MHz
6 V
a
) for ISM bands ranging from 0.15 MHz to
80 MHz
b
)
80% AM at 1 kHz