Recommended separation distances between portable and mobile RF communications equipment
and the ME EQUIPMENT or ME SYSTEM
The thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the thermometer can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the thermometer as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
M
150 KHz to 80 MHz
d = 1,2 √P
80 MHz to 800 MHz
d = 1,2 √P
800 MHz to 2,5 GHz
d = 2,3 √P
0.01 N/A 0.12 0.23
0.1 N/A 0.38 0.73
1 N/A 1.2 2.3
10 N/A 3.8 7.3
100 N/A 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reection from structures, objects and people.
DECLARATION OF EC COMPLIANCE:
Hereby, RADIANT INNOVATION INC declares that the radio equipment type REF. 02009196100000, is in com-
pliance with Directive 2014/53/EU, Directive 93/42/EEC and directive 2011/65/EU (RoHS). The full text of the
EU declaration of conformity is available at the following internet address: http://www.radiantek.com.tw/
download/certifications.asp
This product complies with the Directive 2012/19/EU.
The crossed bin symbol on the appliance indicates that the product, at the end of its life, must
be disposed of separately from domestic waste, either by taking it to a separate waste dispo-
sal site for electric and electronic appliances or by returning it to your dealer when you buy
another similar appli ance. The user is responsible for taking the appliance to a special waste di-
sposal site at the end of its life. If the disused appliance is collected correctly as separate waste,
it can be recycled, treated and disposed of ecologically; this avoids a negative impact on both
the environment and health, and contributes towards the recycling of the product’s materials. For further
information regarding the waste disposal services available, contact your local waste disposal agency or the
shop where you bought the appliance.
COMPLIANCE WITH EU DIRECTIVE 2006/66 CE
The crossed bin symbol on the batteries or product pack indicates that, at the end of their life,
they must not be disposed of as urban refuse. They must be disposed of separately from dome-
stic waste, either by taking them to a separate waste disposal site for batteries or by returning
them to your dealer when you buy similar rechargeable or non-rechargeable batteries. The che-
mical symbols Hg, Cd, Pb, printed under the crossed bin symbol, indicate the type of substance contained in
the batteries: Hg=Mercury, Cd=Cadmium, Pb=Led. The user is responsible for taking the batteries to special
waste disposal sites at the end of their life to allow for their processing and recycling. The correct disposal of
your old toy will help prevent potential negative consequences for the environment and human health and
favour the recycling of the materials of which the product is made. Improper disposal of the product by the
user may harm the environment and human health. For further i nformation regarding the waste disposal
services available, contact your local waste disposal agency or the shop where you bought the appliance.
NOTE:
• The Bluetooth® trademark and logos are registered trademarks of SIG. Inc. and any use thereof by PIKDARE
is under license.
• Google Play and the Google Play logo are Google Inc. trademarks.
• Apple and the Apple logo are trademarks of Apple Inc., registered in the United States and in other coun-
tries. App Store is a service trademark of Apple Inc. registered in the United States and in other countries.
WARNING:
The product is guaranteed against any conformity defect in normal conditions of use as provided for by the
instructions for use. The warranty shall not therefore apply in the case of damages caused by improper use,
wear or accidental events.
For the duration of warranty on conformity defects please refer to the specific provisions of national laws
applicable in the country of purchase, where provided.
In case of problems concerning settings, maintenance or use of the device, contact your local distributor or
the Italian toll-free number 800 900 080.
WARNING: Do not service and/or repair the appliance while it is being used! Any other maintenance/
service activity/operation must be done solely by a PiC Service Centre.
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the
device.)
LOT: 10YYMMXXXX, applicance date of production: the second pair of digits in the lot number identifies the
year, the third pair identies the month (LOT 1016100001; 16=2016; 10=October).
Note: The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy
of temperature measurements, please contact the your local distributor or the Italian toll-free number 800
900 080.
WARNING: No modification of this equipment is allowed
Please read these instructions carefully before use and keep them for future reference
Type BF applied part
WARRANTY
The product is guaranteed against any conformity defect in normal conditions of use as provided for by the
instructions for use. The warranty shall not therefore apply in the case of damages caused by improper use,
wear or accidental events.
For the duration of warranty on conformity defects please refer to the specific provisions of national laws
applicable in the country of purchase, where provided.
In case of problems concerning settings, maintenance or use of the device, contact your local distributor or
the Italian toll-free number 800 900 080. Warning! Do not service and/or repair the appliance.
Guidance and manufacturer’s declaration – electromagnetic emissions
The thermometer intended for use in the electromagnetic environment specified below. The customer or the user
of the thermometer should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The thermometer uses RF energy only for its in-
ternal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The thermometer is suitable for use in all establi-
shments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The thermometer is intended for use in the electromagnetic environment specified below. The customer or the
user of the thermometer should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment –
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
Not applicable
3 V/m
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the thermometer,
including cables, than the recommen-
ded separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance :
d = 1,2 √P
d = 1,2 √P 80 MHz to 800 MHz
d = 2,3 √P 800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manu-
facturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electroma-
gnetic site survey, a should be less than
the compliance level in each frequen-
cy range b.
Interference may occur in the vicinity
of equipment marked with the fol-
lowing symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reection from structures, objects and people.
a
) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the NOME PRODOTTO is
used exceeds the applicable RF compliance level above, the thermometer should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the thermometer.
b
) Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
Guidance and manufacturer’s declaration – electromagnetic immunity
The thermometer is intended for use in the electromagnetic environment specified below. The customer or the
user of the thermometer should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power supply
lines
1 kV for input/output
lines
Not applicable
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Surge
IEC 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
Not applicable
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Not applicable
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the thermo-
meter requires continued operation
during power mains interruptions, it is
recommended that the thermometer
be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
02009196100000
76 009196 100 000
Rev. 01/2017
Distributed dy:
PIKDARE S.r.l.,
Via Saldarini Catelli 10,
22070 - Casnate con Bernate (CO) - Italia
Medical Technology Promedt Consulting GmbH
Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany
Radiant Innovation Inc. Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9th Road, Science-Based Industrial Park,
HsinChu, Taiwan 300.
0120
(93/42/EEC)
(2014/53/EU)