Platypus DAF Jar Tester
Question and Answer IconNeed help?

Do you have a question about the platypus DAF Jar Tester and is the answer not in the manual?

platypus DAF Jar Tester Specifications

General IconGeneral
ModelDAF Jar Tester
Number of Stirrers6
TypeLaboratory Test Equipment
ApplicationDissolved Air Flotation (DAF) simulation and optimization
Capacity2 Liters per jar
Stirrer Speed Range300 RPM
Stirrer MaterialStainless Steel
IlluminationAdjustable LED lighting from below
Power Supply220-240V AC, 50/60Hz

Summary

Introduction to Platypus DAF Jar Tester

Operating Conditions for Platypus DAF Jar Tester

Platypus DAF Saturator Components

Platypus DAF Jar Tester Operating Instructions

DAF Saturator Preparation Procedure

Details the procedure for preparing the DAF Saturator before commencing a test.

Saturator Outlet and Distribution System Preparation

Outlines the preparation of the outlet and distribution system for optimal DAF jar test performance.

Performing the DAF Jar Test

Details the procedure for conducting a DAF jar test, including transferring saturated solution and observing results.

Refilling the Saturator

Instructions on how to refill the saturator, referencing the initial preparation steps.

Calculating Recycle Rate for DAF Tests

Explains the formula and provides an example for calculating the recycle rate during a DAF jar test.

Preparing the Saturator for Storage

Maintenance of the Platypus DAF Jar Tester

Supplementary Jar Testing Parameters