PROFESSIONAL
PROCEDURE
GUIDE
MULTI-TEST A1C SYSTEM
Preparation
1 2 3 4 5 7 986
BEFORE YOU BEGIN COLLECT BLOOD INSERT BLOOD
COLLECTOR
SHAKE INSERT CARTRIDGE DISPENSE SAMPLE
INTO CARTRIDGE
REUSE MONITOR5 MINUTES TO RESULTSPREPARE SAMPLER
Blood Collection Blood Dilution Blood Testing Results
°C °F
Run the test with all
parts of the test kit at
the same temperature
within the specied
range.
If the kit has recently been at high
temperatures (above 82ºF) or in the
refrigerator, keep the kit at room
temperature for at least one hour before
use.
Avoid running the test in direct sunlight,
on hot or cold surfaces, or near sources of
heat or cold.
Quality control materials should be used
to conrm the test kit is working properly.
Refer to the product insert for information
on when to run controls.
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
0628216
Lot Number and Dating Label
Monitor (back)
Pouched
Test Cartridge
Ensure Lot #’s match
Complete
test within
15 minutes.
Sampler
Body
Blood
Collector
Base
(do not remove!)
WAIT TO OPEN
Or
45º
x
x
x
MATCH
CODES
SAMP LE
SAMP LE
SAMP LE
A1
A1
% A1C
A09
LOT 0628216
EXP 21 SEP 14
LOT 0628216
EXP 21 SEP 14
LOT 0628216
EXP 21 SEP 14
MATCH
CODES
% A1C
% A1C
MATCH
CODES
SAMP LE
SAMP LE
SAMP LE
A09
A09
% A1C
% A1C
SAMP LE
SAMP LE
SAMP LE
A09
% A1C
SAMP LE
SAMP LE
SAMP LE
A09
A09
% A1C
MATCH
CODES
Venous Draw
Mix blood
well before
testing
Too little
add more
blood
Not fully inserted
Keep pushing!
* If not, call Customer Service at
1-877-870-5610.
If you cannot resolve an error, please
call Customer Service at
1-877-870-5610.
Do not handle
Monitor again until
test is complete!
Do Codes
Match? *
Monitor and Test Cartridge codes
must match
Ensure Monitor is on level surface
WAIT for SMPL to display
Ready for Sampler
Fully inserted
Just right Too much
wipe away
excess
Collect blood from a slide
Push rmly!
Fingerstick
Shake well 6-8 times
This will mix the blood
with the solution
Push down completely to
dispense diluted sample
Remove quickly
Stand Sampler on table
while preparing Cartridge
“Click” Test Cartridge into place
Use your own
lancet device
to draw blood
Fully insert
Blood Collector into
Sampler Body
Keep
pushing!
Twisting
motion
helps
Gently touch blood
drop to ll
OPEN NOW
Use within 2 minutes.
Remove
Base
Display
counts down
QC test result*
# of tests left
Result
This result cycle remains displayed for
60 minutes or until the next Test Cartridge
is inserted.
THE MONITOR IS REUSABLE
To run another test, use a new
Sampler and Test Cartridge from the
same kit and return to Step 1,
PREPARATION.
ALWAYS MATCH LOT NUMBERS
Use Monitor only with the materials
included in the original kit. The Monitor
will expire after the programmed number
of tests have been run. If another Test
Cartridge is inserted, the Monitor will
display “00 TL.”
* If “QCOK” is not displayed, please see
list of error codes on reverse side.
Discard
Test Cartridge
Save
Monitor
Lot
Code
Product Dating
& Lot ID Label
Lot
Number
91078 B 3/2014
Question and Answer IconNeed help?

Do you have a question about the Polymer Technology Systems A1C Now+ and is the answer not in the manual?

Polymer Technology Systems A1C Now+ Specifications

General IconGeneral
BrandPolymer Technology Systems
ModelA1C Now+
CategoryBlood Glucose Meter
LanguageEnglish

Summary

A1C Now+ Professional Procedure Guide

Preparation and Temperature

Ensure test kit parts are at room temperature before use to maintain accuracy.

Quality Control Materials

Use quality control materials to confirm test kit functionality.

Test Completion Window

Complete the test within 15 minutes of starting the procedure.

Blood Collection and Preparation

Fingerstick Blood Collection

Collect blood from a fingerstick by gently touching the drop to fill the collector.

Blood Sample Volume Check

Ensure correct blood sample volume; add more if too little, wipe excess if too much.

Blood Collector Insertion

Fully insert the Blood Collector into the Sampler Body until it clicks.

Mixing Blood Sample

Shake the device well for 6-8 times to mix blood with the solution.

Cartridge and Sample Handling

Test Cartridge Insertion

"Click" the Test Cartridge into place, ensuring it is properly seated.

Monitor Ready State

Wait for the monitor to display "SMPL" indicating it is ready for the sample.

Code Matching Verification

Ensure Monitor and Test Cartridge codes match; call customer service if they do not.

Dispense Diluted Sample

Push down completely to dispense the diluted sample into the cartridge quickly.

Result and Monitor Reuse

Sample Dispensing Time Limit

Use the dispensed sample within 2 minutes of dispensing it.

Result Countdown

The display counts down the time until the results are available.

Monitor Handling During Test

Do not handle the monitor again until the test is complete.

Monitor Reusability

The A1C Now+ monitor is reusable for subsequent tests.