PROFESSIONAL
PROCEDURE
GUIDE
MULTI-TEST A1C SYSTEM
Preparation
1 2 3 4 5 7 986
BEFORE YOU BEGIN COLLECT BLOOD INSERT BLOOD
COLLECTOR
SHAKE INSERT CARTRIDGE DISPENSE SAMPLE
INTO CARTRIDGE
REUSE MONITOR5 MINUTES TO RESULTSPREPARE SAMPLER
Blood Collection Blood Dilution Blood Testing Results
°C °F
• Run the test with all
parts of the test kit at
the same temperature
within the specied
range.
• If the kit has recently been at high
temperatures (above 82ºF) or in the
refrigerator, keep the kit at room
temperature for at least one hour before
use.
• Avoid running the test in direct sunlight,
on hot or cold surfaces, or near sources of
heat or cold.
• Quality control materials should be used
to conrm the test kit is working properly.
Refer to the product insert for information
on when to run controls.
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
0628216
Lot Number and Dating Label
MATCH
CODES
Venous Draw
Mix blood
well before
testing
Too little
add more
blood
Not fully inserted
Keep pushing!
* If not, call Customer Service at
1-877-870-5610.
If you cannot resolve an error, please
call Customer Service at
1-877-870-5610.
Do not handle
Monitor again until
test is complete!
Do Codes
Match? *
Monitor and Test Cartridge codes
must match
Ensure Monitor is on level surface
WAIT for SMPL to display
Ready for Sampler
Fully inserted
Just right Too much
wipe away
excess
Collect blood from a slide
Push rmly!
Fingerstick
Shake well 6-8 times
This will mix the blood
with the solution
Push down completely to
dispense diluted sample
Remove quickly
Stand Sampler on table
while preparing Cartridge
“Click” Test Cartridge into place
Use your own
lancet device
to draw blood
Fully insert
Blood Collector into
Sampler Body
Keep
pushing!
Twisting
motion
helps
Gently touch blood
drop to ll
OPEN NOW
Use within 2 minutes.
Remove
Base
Display
counts down
QC test result*
# of tests left
Result
This result cycle remains displayed for
60 minutes or until the next Test Cartridge
is inserted.
THE MONITOR IS REUSABLE
To run another test, use a new
Sampler and Test Cartridge from the
same kit and return to Step 1,
“
PREPARATION.”
ALWAYS MATCH LOT NUMBERS
Use Monitor only with the materials
included in the original kit. The Monitor
will expire after the programmed number
of tests have been run. If another Test
Cartridge is inserted, the Monitor will
display “00 TL.”
* If “QCOK” is not displayed, please see
list of error codes on reverse side.
Discard
Test Cartridge
Save
Monitor
Lot
Code
Product Dating
& Lot ID Label
Lot
Number
91078 B 3/2014