the applicable requirements of relevant
electromedical standards, including EN
60601-1-2:2007.
The my5model must emit
electromagnetic energy in order to
perform its intended function. Nearby
electronic equipment may be affected
by this emission. Similarly the my5™
model may itself be affected by
electromagnetic emissions from other
equipment in the vicinity.
The my5model is intended for use
in the electromagnetic environment
specified below. The customer or user of
the my5 model should assure that it is
used in such an environment:
Floors should be wood, concrete or
ceramic tile. If the floor is covered with
synthetic material, the relative humidity
in the room should be at least 30%.
Mains power quality should be that of a
typical domestic, commercial or hospital
environment as appropriate.
The my5model should not be used
during power mains interruptions.
Power frequency magnetic elds
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
The my5™ model is intended for
use in an electronic environment in
which radiated radio-frequency (RF)
disturbances are controlled.
The customer or user of the my5™
model can help prevent electromagnetic
interference by maintaining a minimum
distance between portable and mobile RF
communications equipment (transmitters)
and the my5model as recommended
below, according to the maximum output
power of the communications equipment:
The adequate separation distance
to provide base immunity to RF
disturbances is 1.0 metres.
Retain this user manual for your future
reference.
08