Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro7 is intended for use in the electromagnetic environment specied below. The customer or the user of
Pro7 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or
discharge (ESD) ±8 kV air ±8 kV air ceramic tile. If oor are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical fast ±2 kV for power ±2kV for power Mains power quality should be that of
transient/burst supply lines supply lines a typical commercial or hospital IEC
61000-4-4 environment.
Surge ± 1 kV line(s) t ±1 kV dierential Mains power quality should be that of
IEC 61000-4-5 o line(s) mode a typical commercial or hospital
environment.
Voltage dips, short <5% U
T
<5% U
T
Mains power quality should be that
interruptions and (>95% dip in U
T
) (>95% dip in U
T
) of a typical commercial or hospital
voltage variations for 0.5 cycle for 0.5 cycle environment. If the user of the Pro7
on power supply requires continued operation during
input lines 40% U
T
40% U
T
power mains interruptions, it is IEC
61000-4-11 (60% dip in U
T
) (60% dip in U
T
) recommended that the Pro7 be powered
for 5 cycles for 5 cycles from an uninterruptible power supply or a
battery.
70% U
T
70% U
T
(30% dip in U
T
) (30% dip in U
T
)
for 25 cycles for 25 cycles
<5% U
T
<5% U
T
(>95% dip in U
T
) (>95% dip in U
T
)
for 5 sec for 5 sec
Power frequency
3A/m
3A/m
Power frequency magnetic elds should
(50Hz) magnetic be at levels characteristic of a typical
61000-4-8
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.