Questions and Answers

  • A
    Anne ClarkSep 10, 2025
    What does it mean if 'CAL FAIL' appears on the display of my Precision Medical Oxygen Equipment?
    • A
      Angela AlexanderSep 10, 2025
      If 'CAL FAIL' appears on the display, it could be due to: 1. An improper or wrong oxygen sensor. Try to recalibrate the Oxygen Monitor. 2. An improper air/oxygen source. Check the air/oxygen source. 3. A non-functioning oxygen sensor. Replace it with a new Precision Medical Oxygen Sensor.
  • J
    Joseph GreeneAug 23, 2025
    Why does Precision Medical Oxygen Equipment show a triple pulse alarm?
    • B
      Brittany DodsonAug 23, 2025
      A triple pulse alarm and flashing LED on your Precision Medical Oxygen Equipment indicates that oxygen readings are outside the set high/low alarm limits or there is a loss of air or oxygen supply. You can resolve this by: * Adjusting the high/low alarm settings to be above or below the displayed oxygen value. * Reconnecting the air or oxygen supply.
  • M
    Melissa JohnsonAug 25, 2025
    What to do if Precision Medical PM5900 displays 'CAL FAIL'?
    • A
      Anne RobertsAug 25, 2025
      If the 'CAL FAIL' message appears on the display, it could be due to: * An improper or wrong oxygen sensor: Try to recalibrate the Oxygen Monitor. * An improper air/oxygen source: Check the air/oxygen source. * A non-functioning oxygen sensor: Replace it with a new Precision Medical Oxygen Sensor.
  • D
    Denise LandryAug 27, 2025
    What to do if Precision Medical PM5900 keys are inoperable?
    • O
      ooneillAug 27, 2025
      If the keys on your Precision Medical Oxygen Equipment are inoperable while the power is on, the keypad may be locked. To resolve this, unlock the keypad.
  • D
    David MartinezAug 27, 2025
    How to fix a Precision Medical PM5900 Oxygen Equipment with no display?
    • L
      Lawrence CarterAug 28, 2025
      If there is no display or the LCD screen will not power on, it could be due to: * Dead batteries: Check and replace them with 4 new AA alkaline batteries. * Batteries installed incorrectly: Ensure that the batteries are installed in the proper position (+/-).
  • U
    uleAug 29, 2025
    What to do if Precision Medical PM5900 sensor does not react?
    • W
      William TorresAug 29, 2025
      If the oxygen sensor does not react to changes in oxygen concentration, it could be due to: * Condensation on the oxygen sensor: Remove the condensation. * A non-functioning oxygen sensor: Replace it with a new Precision Medical Oxygen Sensor.
  • C
    carlsonstephenSep 18, 2025
    What to do if the oxygen sensor does not react to changes in oxygen concentration in Precision Medical PM5900?
    • A
      Amanda MccormickSep 18, 2025
      If the oxygen sensor does not react to changes in oxygen concentration, it may be due to condensation on the oxygen sensor. Remove the condensation. Alternatively, the oxygen sensor may be non-functioning, in which case you should replace it with a new Precision Medical Oxygen Sensor.
  • T
    Theodore RobertsSep 12, 2025
    What does it mean when my Precision Medical PM5900 displays 'SERVICE NEED'?
    • E
      Edward WhitakerSep 12, 2025
      If 'SERVICE NEED' appears on the display, it indicates an electronic malfunction. The Oxygen Monitor must be serviced.
  • H
    Heather StoneSep 8, 2025
    Why does my Precision Medical PM5900 show a low battery indicator?
    • D
      Daniel TaylorSep 8, 2025
      If the low battery indicator is displayed with a 30-second chirp alarm, it means the battery voltage is too low. Replace the batteries with four new AA alkaline batteries.
  • Z
    zcoxSep 5, 2025
    What to do if 'Check Sensor' appears on my Precision Medical Oxygen Equipment display?
    • T
      tbartonSep 5, 2025
      If 'Check Sensor' appears on the display with a continuing pulsing alarm, there are several potential causes: * First, ensure that all cable connections are secure and that the locking rings are tight. * Second, the oxygen sensor may not be functioning correctly, in which case you should replace it with a new Precision Medical Oxygen Sensor. * Third, make sure you are using a Precision Medical Oxygen Sensor. If not, attach a new one.

Precision Medical PM5900 Specifications

General IconGeneral
TypeOxygen Concentrator
Flow Rate0.5 - 5 LPM
Oxygen Purity93% ± 3%
Input Frequency50/60 Hz
Operating Temperature41°F to 104°F (5°C to 40°C)
Power SupplyAC
Operating HumidityUp to 95% RH non-condensing