EasyManua.ls Logo

Preston PS-425 - Testing Aircraft Static System; Testing the Aircraft Pitot and Static Systems Simultaneously

Default Icon
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
7. Go to approximately 75% of the maximum airspeed of aircraft indicator (or the
speed specified by the manufacturer). Close the Pressure Control valve and after
allowing a few seconds for stabilization, perform a leak check. The pitot system
should not leak more than two knots per second, plus the leakage previously noted
in the pressure system pre-test.
8. Gently open the Pressure Vent valve to return the pitot pressure to ambient
pressure.
Testing of Aircraft Static System
Note: If the previous pitot system test indicated an excessive leak, do not proceed until
that leak is repaired. Damage to aircraft airspeed indicator could result.
1. Turn on the power to the PS-425.
2. Connect the Vacuum Port to the aircraft static port, and the Pressure Port to the
aircraft pitot port. Set the aircraft altimeter to 29.92 inHg.
3. Close Vacuum Control, Vacuum Vent, Pressure Control, and Pressure Vent
valves. Open Crossbleed valve.
4. Actuate the Vacuum Pump until a desired level of vacuum is seen on the Vacuum
Reservoir gauge (generally -15 to -20 inHG). Actuate the Pressure Pump until the
Pressure Reservoir indicates 15 PSI or more.
5. Gently open the Vacuum Control valve to cause the altitude to climb.
6. DO NOT exceed the climb rate of the aircraft VSI indicator, or the airspeed
indicator limits. The VSI is a very delicate instrument.
7. Climb to the desired altitude(s) per the test to be performed.
8. Close the Vacuum Control valve, and after allowing a few seconds for
stabilization, compare the aircraft altimeter with the test box.
9. When at the appropriate altitude (generally 18,000 feet) perform a leak check. If
the static system shows leakage, be careful to not exceed the VSI rate. Be
prepared to open the Vacuum Control valve if needed, to prevent excessive VSI
indication.
10. Gently open the Pressure Vent valve to return the system pressure to ambient
pressure. Do not exceed the VSI or airspeed indicator limits.
Note: if an altitude lower than ambient pressure is desired, do steps 1-4 above. Gently
open the Pressure Control valve to pressurize the system, thereby causing the
altimeter to decrease in altitude. Open the Vacuum Vent valve to return to ambient
pressure.
Testing the Aircraft Pitot and Static systems simultaneously
Note: First ensure pitot system does not leak by performing Pitot system test on page 14.
1. Turn on the power to the PS-425.
2. Connect the Vacuum Port to the aircraft static port, and the Pressure Port to the
aircraft pitot port. Set the aircraft altimeter to 29.92 inHg (1013.3 mB).