(111) PTSA
0 - 1000 µg/l (PTSA)
Do ONLY use the cuvette which has been used to do calibration
for this parameter!
Use 10ml Pipette to always properly dose exactly 10ml!
1) Fill sample water into the same cuvette up to the 10ml level.
2) Insert the cuvette WITHOUT LID into the PrimeLab sampling chamber.
Align the arrow on the front of the cuvette.
3) Place PTSA-Adapter on top of the open cuvette which already is placed
into the PrimeLab measurement chamber.
4) Press TEST.
5) The determined result is immediately displayed. Press button 3 to
convert the result to different units (*). Press button 4 to save the result
together with date and time.
10 ml
Name on device: 111-PTSA-Ad
24 mm
4
Adapter
PLSp-PIP10ml PLSp-PIP10ml-tipsPLSp-ADP-PTSA
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
3
4
PLSp-RefPTSA
Only for calibration!