Powder
Continued...
11) After the lapse of a 2 minute countdown the determined result is
displayed. Press button 3 to convert this into various measurement units
(*).
12) Press button 4 to save the result together with date and time in the
device.
3
4
Note(s)
-> (*) Conversion to Si
-> The temperature of the water sample must be between 15°C and 25°C to
ensure precise measurements.
-> Sulfide in the water sample will influence the measurement result.
-> Larger amounts of iron falsify the measurement result.
-> Phosphate content in the water higher that 60 mg/l will falsify the
measurement result.
Name on device: 50-Silica-HR-pow
(50) Silica (HR)
0 - 100 mg/l (SiO )
2
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.