Pro'sKit MT-1236 - Diode Test Accuracy; DC Current Measurement Accuracy; AC Current Measurement Accuracy

Pro'sKit MT-1236
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.2.7 Diode
Functions
Testing conditions
Diode test
Forward DC current: approximately 1.2mA; Open-circuit
voltage: approximately 3.2V.
The display unit shows the approximate value of the diode's
forward voltage drop.
The buzzer beeps when the resistance is less than 30
Overload protection: 250V DC/AC
4.2.8 DC current
Measuring range
Resolution
Accuracy
60mA
0.01mA
±(1.2% Reading + 5 digits)
600mA
0.1mA
10.00A
10mA
±(3% Reading + 5 digits)
Overload protection:
Protective tube for mA measuring range (F600mA/250V) ;
Protective tube for 10A measuring range (F10A/250V) .
Maximal input voltage: mA position: 600mA DC/AC (valid value);
10A position: 10A DC/AC (valid value)
When the measured current is over 5A, the duration of continuous measurement
shall not be over 10 seconds. The current measurement shall be carried out 1
minute after the completion of previous measurement.
4.2.9 AC current
Measuring range
Resolution
Accuracy
60mA
0.01mA
±(1.5% Reading + 5 digits)
600mA
0.1mA
10.00A
10mA
±(3% Reading + 5 digits)
Overload protection:
Protective tube for mA measuring range (F600mA/250V) ;
Protective tube for 10A measuring range (F10A/250V) .
Maximal input voltage: mA position: 600mA DC/AC (valid value);
10A position: 10A DC/AC (valid value)
When the measured current is over 5A, the duration of continuous
measurement shall not be over 10 seconds. The current measurement shall
be carried out 1 minute after the completion of previous measurement.
Frequency response: 40Hz-1kHz True RMS