OVERVIEW
Read the Before You Begin section on page 11 of the ProFiller 3700/3800 Manual v 1.2.1 before using this machine.
A. ORIENTATION OF OUTER CAPSULES
Orient the capsules as shown in the ProFiller 3700/3800 Manual v 1.2.1 Operating Instructions for Orientation, page 12 steps 1-9 using the Orienter and
Caps Tray for the outer capsule size.
B. SEPARATION OF CAPSULE CAPS AND BODIES
1. Make sure Cam Handle is at 0° on cam
indicator. Place loaded Caps Tray with Push
Plate onto Filler.
2. Pull Cam Handle towards you until you feel
a noticeable resistance, securing capsule
bodies in Filler.
3. To separate size AAel capsules, see Inset
1 on next page. For other capsule sizes,
slowly push down Lifting Handle just until
capsules separate. As soon as you begin to
feel resistance, slowly release the Lifting
Handle. This resistance occurs as the
Bottom Plate begins to lift the capsule bodies.
Continuing past this point will cause the
capsule bodies to be lifted out of the Filler.
See Capsule Separation Tip below.
Capsule Separation Tip
If more than 1 or 2 caps fail to separate from bodies in step 3 when filling sizes AAA AA, AAel or A, or if more than 8 to 10 fail to separate for sizes B or D, Cam Handle
should be rotated to a higher degree on Cam indicator in step B2. When rotating Cam Handle, there should be enough pressure on capsule bodies to hold them inside
Filler, but not so much that capsules are squeezed to an oval shape.