How to reduce high luminescent cross talk in Promega Laboratory Equipment?
- ZzochoaAug 16, 2025
Use high-quality white or black opaque plate. If using 384-well plates, confirm that the 384-well aperture has been installed.
How to reduce high luminescent cross talk in Promega Laboratory Equipment?
Use high-quality white or black opaque plate. If using 384-well plates, confirm that the 384-well aperture has been installed.
What to do if the Promega Laboratory Equipment shows 'door opened'?
If the instrument door was opened during a run or something prevented the instrument door from fully closing as the Multiwell Plate Sample Tray was retracting, restart the run or check the instrument door for any obstructions.
Category | Microplate Reader |
---|---|
Detection Modes | Luminescence |
Light Source | Xenon flash lamp |
Detection Technology | Photomultiplier Tube (PMT) |
Interface | USB |
Power Requirements | 100-240 V, 50/60 Hz |