8. The resistance range posions where the device under test indicates PASS are those resistance
values accepted by the wrist strap tester.
a. Resistance posions below the PASS range should all be rejected by the device under test
as too low.
b. Resistance posions above the PASS range should all be rejected by the device under test
as too high.
B. Conrming Foot Wear Tester Resistance Performance Range
1. Plug the two supplied test leads into the PAR-809C receptacles.
2. Aach an alligator clip to one PAR-809C test lead. Aach the alligator clip to the Foot Wear test
plate.
6
Rev. B / January 2012
PAR-809C Variable Resistance Reference