8. The resistance range posions where the device under test indicates PASS are those resistance
values accepted by the wrist strap tester.
a. Resistance posions below the PASS range should all be rejected by the device under test
as too low.
b. Resistance posions above the PASS range should all be rejected by the device under test
as too high.
B. Conrming Foot Wear Tester Resistance Performance Range
1. Plug the two supplied test leads into the PAR-809C receptacles.
2. Aach an alligator clip to one PAR-809C test lead. Aach the alligator clip to the Foot Wear test
plate.
6
Rev. B / January 2012
PAR-809C Variable Resistance Reference