Rev. C / January 2012
PRS-812 Resistance Meter
(3) MANUAL: displays data in exponenal format 1.0EXX
(4) MANUAL: displays data in Ω, KΩ, MΩ, GΩ and TΩ
In , the operator selects resistance range in single decade
increments. Test voltage may be selected by the operator, or allowed
to funcon automacally based on resistance range selected. The
operator determines electricaon period in seconds (SEC) using the
displayed mer in the center of the LCD.
(5) AUTO-MANUAL: displays data in exponenal format 1.0EXX
(6) AUTO-MANUAL: displays data in Ω, KΩ, MΩ, GΩ and TΩ
In the operator may select the inial resistance range in
decade increments. Once set, the PRS-812 starts the measurement process
from the preset resistance decade, rather than re-zeroing itself for each
measurement. This feature saves measurement cycle me and extends
baery life. In this mode the instrument automacally controls test voltage,
range and measurement electricaon period.