D909.131.00/02 en 2012-07 Ext/Sa
User Manual
NOMEX
NOMEX Multimeter T11049 Firmware Version 1.1 or higher
NOMEX Software S030008 Version 1.1 or higher
Contents
Operating Manual
Technical Manual
Service Manual

Table of Contents

Questions and Answers

  • J
    josephbakerAug 16, 2025
    What to do if the dose rate is too high on my PTW NOMEX L981815 Multimeter?
    • W
      william76Aug 17, 2025
      If the PTW Multimeter detects a dose rate that is too high because the electronic measuring system is out of range, you should reduce the dose rate on the X-ray unit, increase the distance, or reduce the tube voltage or tube current.
  • D
    deanna21Aug 19, 2025
    What to do if the dose rate is too low on my PTW NOMEX L981815?
    • G
      grahamryanAug 19, 2025
      If the PTW Multimeter displays a low dose rate because the electronic measuring system is out of range, you can increase the dose rate on the X-ray unit, reduce the distance, or increase the tube voltage or tube current.
  • D
    Dennis WeissAug 22, 2025
    How to improve reduced dose accuracy on PTW NOMEX L981815?
    • D
      David CruzAug 22, 2025
      If your PTW Multimeter indicates reduced dose accuracy due to a measuring error exceeding 5%, try repeating the measurement. You can also increase the dose rate on the X-ray unit, reduce the distance, or manually set the total filtration in the Measurement menu.
  • K
    Kimberly GrahamAug 25, 2025
    How to fix tube voltage too high error on PTW NOMEX L981815?
    • E
      Elizabeth VasquezAug 26, 2025
      If your PTW Multimeter shows that the tube voltage is too high, it is advised to reduce the tube voltage or select kV auto-ranging.
  • D
    Daniel JohnsonAug 27, 2025
    How to fix tube voltage too low error on PTW Multimeter?
    • M
      martintinaAug 27, 2025
      If your PTW Multimeter shows that the tube voltage is too low, you should increase the tube voltage or select kV auto-ranging.
  • T
    Tracy EatonAug 30, 2025
    What to do if my PTW NOMEX L981815 Multimeter displays a corrupted device dataset?
    • C
      Christopher WilliamsAug 30, 2025
      If the PTW Multimeter identifies a corrupted device data set, confirm with the OK button to reset the device data set to the default values.
  • G
    gloveralyssaSep 2, 2025
    What to do if PTW NOMEX L981815 detects short exposure time?
    • J
      james82Sep 2, 2025
      If your PTW Multimeter detects a short exposure time, repeat the measurement.
  • V
    Victoria JimenezSep 4, 2025
    Why does my PTW NOMEX L981815 Multimeter display 'Low dose rate detected'?
    • A
      Alexis PetersonSep 5, 2025
      If the PTW Multimeter displays 'Low dose rate detected' because NOMEX is in a high measuring range, repeat the measurement.
  • E
    Elijah TaylorSep 7, 2025
    What to do if the exposure time is too short on my PTW Multimeter?
    • T
      Tara MaySep 7, 2025
      If your PTW Multimeter indicates that the exposure time is too short, you should reduce the start delay or increase the exposure time.
  • D
    Dr. Destiny GreenSep 10, 2025
    How to resolve 'Unknown radiation quality' on PTW NOMEX L981815?
    • N
      Nicole ChapmanSep 10, 2025
      If your PTW Multimeter displays 'Unknown radiation quality' because the radiation quality cannot be determined, manually set the total filtration in the Measurement menu.