The QinFLOW Warrior lite Blood and Fluid Warmer, Model QiF-03, is a battery-powered, lightweight, and portable system designed for warming blood, blood products, and IV fluids to help prevent hypothermia in medical settings. It is intended for use by qualified healthcare professionals.
Function Description:
The Warrior lite system is a blood and fluid warmer that operates using a Base Unit (controller) and a rechargeable battery. It connects to a Compact Disposable Unit (CDU) which contains a heat exchanger for warming fluids. The system is designed for both prehospital and hospital critical care transports, as well as surgical and intensive care units. It can warm blood, blood products, and intravenous fluids to a set point of 38 °C (100.4 °F). The system provides visual LED panel indications for power, battery level, and operational status, including warnings for troubleshooting, low battery, and critical battery levels.
Important Technical Specifications:
- Power Source: 18 VDC from a fully charged battery.
- Nominal Input Voltage: 26 A.
- Battery Life Expectancy:
- Lite battery: Up to 250 charging cycles.
- Extra Power battery: Up to 350 charging cycles.
- Charger Input:
- Lite battery charger: 100-240 VAC, 50-60 Hz, Max 1.0 A.
- Extra Power battery charger: 100-240 VAC, 50-60 Hz, Max 2.0 A.
- Defibrillation Proof: CF Applied Part.
- Weight:
- Base Unit with lite battery: Approximately 800 g (1.76 lb).
- Base Unit with Extra Power battery: Approximately 1088 g (2.39 lb).
- Compact Disposable Unit: Approximately 117 g (0.26 lb).
- Dimensions (H x W x L):
- Base Unit with lite battery: Approximately 84×88×115 mm (3.30×3.46×4.52 in).
- Base Unit with Extra Power battery: Approximately 105×88×115 mm (4.13×3.46×4.52 in).
- Compact Disposable Unit: Approximately 72.2×68.5×117.5 mm (2.84×2.70×4.63 in).
- Environmental Conditions:
- Storage & Transport: -20 °C to 70 °C & 93% RH (-4 °F to 158 °F & 93% RH).
- Operation Temperature & Humidity: -5 °C to 50 °C & 90% RH (23 °F to 122 °F & 90% RH).
- Transient Operating Conditions: -20 °C to 50 °C & 90% RH.
- Atmospheric Pressure/Altitude: 549 to 1,060 hPa (-400 to 4,572 meters).
- CDU Shelf Life: 3 years.
- Base Unit Service Life: 5 years.
- Ingress Protection: IP56 (Water and particles ingress rate).
- Maximum Delivery Rate:
- Lite Battery: Up to 170 ml/min.
- Extra Power Battery: Up to 180 ml/min.
- Temperature Set Point: 38 °C (100.4 °F).
- Compliance: IEC 60601-1-12 certified, MIL-STD 461G RE102 & RS103 compliant.
Usage Features:
- Portability: The Warrior lite is designed for space and weight constrained rescue gears, making it suitable for mobile medical environments.
- Battery Options: Available with a standard "lite" battery or an "Extra Power" battery for extended operational range. The Warrior Hybrid SKU offers an AC power supply module for continuous operation.
- Ease of Connection: The battery attaches securely to the back of the Base Unit via latches. The CDU connects to the Base Unit via a standard IV / blood administration set. An Extension Cable can be used to increase flexibility.
- Intuitive Interface: The Base Unit features an ON/OFF switch and LED indicators for power, battery level, and system status.
- Fluid Warming Capacity: The system can warm up to 1.3 liters of parenterally administrated blood/fluids with a lite battery or up to 3 liters with an Extra Power battery, from an inlet temperature of 4 °C (39.2 °F) to 38 °C (100.4 °F).
- Safety Features: Includes warnings for incorrect battery insertion, low battery, and system malfunctions. The CDU is a single-use, sterile, and ready-for-use item.
- Accessories: Optional accessories include a Mounting Unit (for connectivity to standard pole, rail, or stretcher), Extension Cable, and a Soft Carrying Bag that can hold the core system and additional accessories.
Maintenance Features:
- Battery Charging: The system uses designated chargers for the lite and Extra Power batteries. A green LED on the charger indicates full capacity. Charging times vary, with a fully depleted lite battery taking approximately 3 hours and an Extra Power battery taking approximately 4 hours.
- Battery Storage: Store fully charged batteries. For long-term storage, recharge the battery every 4-6 months.
- Cleaning and Disinfection: The Warrior lite Base Unit, battery, Extension Cable, and Mount Accessory are supplied non-sterile and should be surface cleaned/disinfected after each patient use. Cleaning procedures involve using a damp cloth with 70% medical grade alcohol solution or medical grade wipe, followed by a soft bristle brush with 70% medical grade alcohol solution for grooves.
- CDU Handling: The Compact Disposable Unit is for single use only and must not be reused or re-sterilized. It should be discarded after use.
- Troubleshooting: The manual provides detailed troubleshooting steps based on LED indications, covering issues like system activation, reaching set temperature, CDU connection, fluid/blood temperature, battery level, and communication problems.
- Warranty: The Warrior lite is manufactured by Quality in Flow Ltd. and is warranted to be free from manufacturer defects for one year from the date of purchase.
- Support: Contact information for support, service, and subsequent order information is provided.