10
5.3.2.13 Press Program and Down-arrow at the same time:
This enters the setpoint-programming mode.
Display toggles between [SP_1] and [50000]. 18.xxx for the new model
5.3.2.14 Press the Program button repeatedly until the
Display toggles between [SP_5] and [30000],10000 for the new model.
Change SP_5 to 1.000
Note: If SP_5 is unavailable, perform the next step (SPC_5) then come back
to this step; Ensure SPC_5 is set to [000].
You should now be able to set up SP_5 to 1.000. Continue and perform the
next step before locking the meter.
5.3.2.15 Press Program button repeatedly until the
Display toggles between [SPC_5] and [000].
Change SPC_5 to 46.
5.3.2.16 Check all the setting before lock
STEP B LOCK CODE SETUP:
5.3.2.17 Loosen the meter from cabinet.
5.3.2.18 Remove black face meter bezel.
5.3.2.18 Set both DIP Switches from OFF to ON position.
5.3.2.20 Reinstall the bezel and secure the meter.