EasyManuals Logo

QUANTACHROME INSTRUMENTS MULTIPYCNOMETER User Manual

Default Icon
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
31
B.3.5. TOTALc IS THE SUM OF FIVE CONSECUTIVE cP READINGS. ENTER THE
SUM AS TOTALc IN SECTION B.3.3.
B.3.6. c IS THE AVERAGE OF FIVE CONSECUTIVE cP VALUES.
B.3.7. VCd SMALL’S dP1 AND dP2 ARE THE PRESSURES BEFORE AND AFTER
EXPANSION RESPECTIVELY.
# dP1
÷
dP2
=
dP
1
÷
=
2
÷
=
3
÷
=
4
÷
=
5
÷
=
d = TOTALd ÷ 5 =
B.3.8. COLUMN dP LISTS FIVE CONSECUTIVE PASSES OF ITEMS THE dP RATIO
MUST FALL WITHIN +/-0.0005 OF EACH OTHER BEFORE THEIR SUM CAN BE
AVERAGED.
B.3.9. TOTALd IS THE SUM OF FIVE CONSECUTIVE dP READINGS. ENTER THE
SUM AS TOTALd IN SECTION B.3.7.
B.3.10. d IS THE AVERAGE OF FIVE CONSECUTIVE dP VALUES.
VC large VC LARGE
B.3.11. VR SMALL = ( c - 1 ) = ( )
(d - 1 )
B.3.12. VC SMALL = VR SMALL ×