B.4.7. VCf MICRO’S fP1 AND fP2 ARE THE PRESSURES BEFORE AND AFTER
EXPANSION (SECTION 5.5.9.), RESPECTIVELY.
# fP1
÷
fP2
=
fP
1
÷
=
2
÷
=
3
÷
=
4
÷
=
5
÷
=
f = TOTALf ÷ 5 =
B.4.8. COLUMN fP LISTS FIVE CONSECUTIVE PASSES OF ITEMS IN SECTION 5.5.9.
THE fP RATIO MUST FALL WITHIN +/-0.0005 OF EACH OTHER BEFORE THEIR SUM
CAN BE AVERAGED.
B.4.9. TOTALf IS THE SUM OF FIVE CONSECUTIVE fP READINGS. ENTER THE SUM
AS TOTALf IN SECTION B.4.7.
B.4.10. f IS THE AVERAGE OF FIVE CONSECUTIVE fP VALUES.
2.145 cc ( SPHERE ) 2.145 cc
B.4.11. VR MICRO = ( e - f ) = ( ) - ( )
(e - 1 )
B.4.12. VC MICRO = VR MICRO ×