Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device’s warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers.
1st Day
Read the User Guide carefully before
use and store it in a safe place for future
reference
Before every use
Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
Make sure all cables are securely plugged in
before use in the shower
Inspect all cables for cracks or signs for
worn-out parts. Never use a product which
appears to be substandard or faulty. Contact
your local dealer for service
Disinfect the upholstery. please check the
disinfection section in this User Guide for
more information
Inspect the detachable parts. If required,
disinfect the parts with a 70% disinfectant
solution. It is recommended to wipe of any
residue and dirt using a cloth wrung in warm
water and a mild detergent/soap without
chlorine. Let the detachable parts dry before
disinfection
After every use
Wipe the product dry
Daily
Visual inspection. Repair or replace any
damaged or worn-out parts
Ensure that all fasteners are in place and
tted correctly
Check all ttings and buckles for signs of
worn out parts
Be aware of any MAX mark indications
before adjustment
Monthly
Use a cloth with warm water and a mild
detergent to clean the product
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
Inspect the piston rods are greased, extract
the piston rods to check grease
Warewashing machine
This product can be washed at:
60°C in 10 minutes
85°C in 3 minutes
...using a mild detergent in a washing machine
designed for medical devices. Use the machine’s
drying function to dry the product. Clean the
accessories separately.
Before washing
Avoid degreasing of the piston rods, the
actuators should be retracted to minimum
stroke and without load before washing.
Remove all accessories, unplug all cables
and make sure to dismount the battery,
control-box and handset. Leave the
actuators with the cables mounted on the
product (see page 22)
After washing
Ensure correct placement of cables inserted
into the control box when reinstalling the
power unit system (see page 23)
Check that the handset functionalities works
correct
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer’s
instructions and demonstrably recorded.