Page 7
8. Display will read:
Adjust Flow For then Enter Low Flow
Low Flow Rate Rate: 0.0 CFM
Reduce flow to low point of calibration range. Enter rate using keypad and then
press “ENTER”. Keep flow constant for 10 seconds.
9. Display will read:
Verify Linearity
Rate: 0.0 CFM
Adjust flow to various points within the calibration range to verify calibration
accuracy. Press ENTER to accept calibration or CLEAR to reject. At this point
the motor will shut off.
NOTE: If display reads “BAD CALIBRATION”, or “TURBINE FAILURE”
then inaccurate flow data has been entered, or the turbine speed sensor requires
service.
10. Display will read:
Default Run Mode
0=Vol, 1=Time:
Select mode of operation. Select “0” for TOTAL VOLUME, select “1” for
TOTAL TIME.
NOTE: During normal operation, both volume and time are displayed.
11. Display will read:
Enter Target or Enter Target
Time: 00:00 Volume 0 ft3 or 0 liters or 0 m3
Enter default sample time or volume using the keypad.