Mica window density | 1.5 - 2.0 mg/cm² |
---|---|
Type | Geiger Counter |
Measurement Range | 0.001 - 100 mR/hr |
Energy Range (Beta) | 0.16 MeV minimum |
Accuracy | ±15% of reading |
Display | Backlit LCD |
Audio | Adjustable audio output |
Outputs | RS-232 data output |
Power Supply | Two AA alkaline batteries |
Operating Temperature | -10°C to 50°C (14°F to 122°F) |
Dimensions | 150 x 80 x 30 mm |
Essential safety guidelines and handling instructions for the Area Monitor to maintain its condition.
Initial device setup, including detector connection and power supply connection.
Method for responding to and dismissing active alarms on the device.
Describes the primary display interface showing radiation rate, units, and power status.
Procedure for configuring the specific thresholds for Low, High, and Fail alarms.
Guide to downloading and installing the software for managing Area Monitors.
Details on accessing the screen to adjust calibration parameters and settings.
General guidance on configuring the system to send alerts via email or text.
Detailed steps for configuring alert rules and email recipients within the software.
Steps to configure hardware alert rules, including selecting alarm conditions and recipients.
Method to access main calibration settings menu via the device's recessed button.
Guide to adjusting the high voltage setting for the detector.
Instructions for setting the minimum pulse size required for a count.
Guide to setting the conversion factor for count rates to dose rates.
How to configure dead time correction to account for missed counts.
A numbered list outlining the recommended steps for device setup and calibration.
Technical specifications for key calibration parameters: High Voltage, Threshold, Dead Time, Cal Constant.