2.
TCM3
TCM3
and
TCM30
Service
Manual
General
Information,
Continued
Specifications
(continued)
2-10
Monitor
pCO,
channel
Range:
5
-
200
mmHg
or
0.7
-
26.7 kPa.
Accuracy:
Better
than
+
1
mmHg/0.1
kPa
0.1
%
of
reading
at
25
°C.
Electrode
Heat
Range:
0
-
600
mW
+
10
%
of
reading.
Temperature
Settings
37.0
°C,
41.0
°C,
41.5
°C,
42.0
°C,
42.5
°C,
43.0
°C,
43.5
°C,
44.0
°C,
44.5
°C,
45.0
°C.
Accuracy:
Better
than
+
0.1
°C
excluding
sensor.
Alarms
Audio
visual.
The
acoustic
alarm
will
be
activated
after
a
delay
of
5
seconds
when
an
alarm
state
has
occurred.
pO,
alarm
limits:
(selected
by user)
LOW:
0-99
mmHg/0
-
9.9
kPa
HIGH:
0
-
800
mmHg/0
-
99.9
kPa
pCO,
alarm
limits:
(selected
by
user)
LOW:
0-99
mmHg/0
-
9.9
kPa
HIGH:
0
-
200
mmHg/0
-
26.7
kPa
Temperature:
+
0.3
°C
deviation
from
preset
value
wili
cause
alarm.
Battery
low:
Alarm
goes
on
before
the
battery
is
empty
Leakage
Currents
Enclosure
Leakage
Current:
Max.
100
uA
Patient
Leakage
Current:
Max.
10
uA
Continued
on
next
page
199808-1