TCM3
and
TCM30
Service
Manual
Error
Codes
Introduction
Soft
Errors
199808-1
2.
TCM3
The
error
codes
are
divided
into
two
categories.
“Soft
Errors”
are
displayed
in
the
aux.
display
(flashing)
and
an
acoustic
beep
is
heard.
“Hard
Errors”
are
displayed
in
the
pO./pCO,
displays
accompanied
by
a
continuous
acoustic
alarm
for
five
seconds,
Soft
Errors
are
numbered
from
0
to
5
and
are
caused
by
incorrect
operation
or
defective
accessories,
e.g.
electrodes
or
electrode
membranes.
Hard
Errors
are
numbered
from
10
to
16
and
are
normally
related
to
electronic
disturbances
or
a
defective
Monitor.
However,
Error
10
and
11
may
be
caused
by
a
defective
electrode
(temperature
sensors).
The
table
lists
the
error
codes
as
well
as
the
possible
causes
and
Service
actions.
Error
Cause
Service
Action
Code
0
RAM
has
lost
its
content
due
to,
|
Check
setting
in
service
program,
e.g.;
the
internal
battery
is
step
4
(mmHg/kPa
and
in-vivo
completely
exhausted
or
cal.).
Reprogram
the
Setup
and
disconnected.
AUX
programs.
1
The
electrode
has
not
reached
the
|
Re-calibrate.
If
error
persists,
pre-set
temp.
within
3
minutes.
teplace
the
electrode.
Defective
electrode.
Replace
the
electrode.
2
Electrode
signal
range
not
-Recalibrate.
Remembrane
and
acceptable.
then
recalibrate.
Incorrect
preset
calibration.
Check
the
value
and
recalibrate.
The
membrane
has
not
been
Remembrane
and
then
recalibrate.
changed
regularly.
Unsuccessful
membraning.
Remembrane.
Poor
supply
of
calibration
gas.
Check
the
gas
source
and
connection.
Defective
electrode.
Replace
the
electrode.
3
Electrode
stability
not
acceptable.
|
Recalibrate.
Remembrane
and
then
recalibrate.
Membrane
has
not
been
changed
|
Remembrane
and
recalibrate.
regularly.
Unsuccessful
membraning.
Remembrane.
Poor
supply
of
calibration
gas.
Check
the
gas
source
and
connection.
Defective
electrode.
Replace
the
electrode.