EasyManua.ls Logo

R&S FSH6 - 3 GPP FDD Code Domain Power Specs; 3 GPP FDD BTS;Node B Measurement Details

R&S FSH6
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
¸FSH 3 ¸FSH 6
3GPP FDD code domain power BTS/Node B measurement (only with R&S®FSH-K4 1300.7633.02 and ¸FSH 3 model .23)
6)
Frequency range 10 MHz to 3 GHz
Carrier frequency uncertainty (test case 6.3 in line with
3GPP 25.141)
Measurement range ±1 kHz
Measurement uncertainty
SNR > 30 dB
< 50 Hz + Δf
ref
7)
(σ = 20 Hz)
Total power SNR > 30 dB (test case 6.2.1 in line with
3GPP 25.141)
Measurement range
frequency > 1 MHz
+20 °C to +30 °C
–60 dBm < P
total
< 20 dBm
Measurement uncertainty
–40 dBm < P
total
< 20 dBm
P
REF_LEV
30 dB < P
total
< P
REF_LEV
+ 3 dB
±1.5 dB, typ. 0.5 dB
CPICH power SNR > 30 dB (test case 6.2.2 in line with
3GPP 25.141)
Measurement range
–40 dBm < P
total
< 20 dBm P
total
20 dB < P
CPICH
< P
total
Measurement uncertainty
– P
total
–20 dBm < P
CPICH
< P
total
±1.5 dB, typ. 0.5 dB
P-CCPCH power SNR > 30 dB
Measurement range
–40 dBm < P
total
< 20 dBm P
total
40 dB < P
PCCPCH
< P
total
Measurement uncertainty
P
total
–20 dBm < P
PCCPCH
< P
total
±1.5 dB, typ. 0.5 dB
PSCH/SSCH power SNR > 30 dB
Measurement range
–40 dBm < P
total
< 20 dBm P
total
30 dB < P
SCH
< P
total
Measurement uncertainty
P
total
20 dBm < P
PSCH
< P
total
±2.5 dB, typ. 1.5 dB
Symbol EVM
Measurement range 3 % < EVM
symbol
< 25 %
Measurement uncertainty 3 % < EVM
symbol
< 10 % typ. ±2.5 %
10 % < EVM
symbol
< 20 % typ. ±3 %
Residual EVM
symbol
typ. 3 %
Handheld Spectrum Analyzer ¸FSH20
6)
As of serial no. 103500.
7)
Δf
ref
= uncertainty of reference frequency.

Related product manuals