¸FSH 3 ¸FSH 6
3GPP FDD code domain power BTS/Node B measurement (only with R&S®FSH-K4 1300.7633.02 and ¸FSH 3 model .23)
6)
Frequency range 10 MHz to 3 GHz
–
Carrier frequency uncertainty (test case 6.3 in line with 
3GPP 25.141)
–
  
Measurement range ±1 kHz
–
  Measurement uncertainty
SNR > 30 dB
< 50 Hz + Δf
ref
7)
 (σ = 20 Hz)
–
Total power SNR > 30 dB (test case 6.2.1 in line with  
3GPP 25.141)
–
  Measurement range
frequency > 1 MHz 
+20 °C to +30 °C
–60 dBm < P
total
 < 20 dBm
–
  Measurement uncertainty
–40 dBm < P
total
 < 20 dBm 
P
REF_LEV
 – 30 dB < P
total  
< P
REF_LEV
+ 3 dB 
±1.5 dB, typ. 0.5 dB
–
CPICH power SNR > 30 dB (test case 6.2.2 in line with  
3GPP 25.141)
–
  Measurement range
–40 dBm < P
total
 < 20 dBm P
total
 – 20 dB < P
CPICH
 < P
total
–
  Measurement uncertainty
– P
total
 –20 dBm < P
CPICH
 < P
total
±1.5 dB, typ. 0.5 dB
–
P-CCPCH power SNR > 30 dB
  Measurement range
–40 dBm < P
total
 < 20 dBm P
total
 – 40 dB < P
PCCPCH
 < P
total
–
  Measurement uncertainty
P
total
 –20 dBm < P
PCCPCH
 < P
total
±1.5 dB, typ. 0.5 dB
–
PSCH/SSCH power SNR > 30 dB
–
  Measurement range
–40 dBm < P
total
 < 20 dBm P
total
 – 30 dB < P
SCH
 < P
total
–
  Measurement uncertainty
P
total
 –20 dBm < P
PSCH
 < P
total
±2.5 dB, typ. 1.5 dB
–
Symbol EVM
  Measurement range 3 % < EVM
symbol
 < 25 %
–
  Measurement uncertainty 3 % < EVM
symbol
 < 10 % typ. ±2.5 % 
–
10 % < EVM
symbol
 < 20 % typ. ±3 % 
  Residual EVM
symbol
typ. 3 % 
Handheld Spectrum Analyzer ¸FSH20
6)
  As of serial no. 103500.
7)
  Δf
ref
 = uncertainty of reference frequency.