Ranger Fluid Warming Systems
English
15
Specifications
Physical characteristics
 
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long;
wt.: 7 lb. 7 oz. (3.4 kg)

Classified under IEC 601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B,
Ordinary Equipment. Not suitable for use in the presence of flammable anesthetic mixtures with air or
with oxygen or nitrous oxide. Classified by Underwriters Laboratories Inc. with respect to electric shock,
fire, and mechanical hazards only, in accordance with UL 60601-1 and in accordance with Canadian/CSA
C22.2 No. 601.1.
Electrical characteristics Temperature characteristics
    
100-120 or 220-240 VAC 41°C
  - 
100-120 VAC, 50/60 Hz Model 245: 43°C
220-240 VAC, 50 Hz Model 247: 48°C
   - 
900 W 33°C
 - 
2 x 10 Amp for 100 – 120 VAC Model 245: 44°C
2 x 6.3 Amp for 220 – 240 VAC Model 247: 50°C
Leakage current
Meets leakage current requirements in accordance with UL 60601-1 and IEC 601-1.
Environmental conditions
  
15° to 40°C (59° to 104°F)
  
-20° to 45°C (-4° to 113°F)
 
10 to 85% RH, noncondensing
  
50 kPa to 106 kPa