2.
1.
3.
4.
Required calibration cards: 1 x pink calibration card, 10 x grey calibration cards,
10 x white calibration cards
CALIBRATION INSTRUCTIONS rapidcount T 575
EN
ratiotec GmbH & Co. KG | Tel.: +49 201 361 490 | Fax: +49 201 361 49 26 | Email: info@ratio-tec.de
Connect the power adapter with the device.
Turn on the device. Then press the “CLEAR“
button times. A window with „Input Pass-
word“ opens on the display . Please enter a 0
and conrm it with the “START“ button.
Now the “Maintenance Menu“ opens. Select
“Sensor Adjust“ and conrm it with the
“START“ button. Afterwards please press the
“REPORT“ button to get to “Static Adjust“.
Conrm it again with the “START“ button.
Open the cover of the device so that the feed
rollers are visible. Place the pink calibration
card centrally under the feed roller. Then close
the cover and press the “START“ button. The
calibration card is automatically ejected after
the calibration.
Press the “REPORT“ button to get to “Image
Dynamic Adjust“. Conrm it with the “START“
button. Place all gray calibration cards on the
upper feed. Repeat this process until “Press
START key“ is shown.
pink
calibration card!
grey
calibration cards!