English Instruction Manual DIGIMON-SE / DIGIMON4
10
displays the difference "K" (F) between the temperature according to the vapour
pressure table (to, tc) and the measured temperature (t1, t2) of the external
thermocouples (9) or the clamp thermocouples (10).
Vacuum display
Press the button to bring up the vacuum display.
Vacuum display values
When evacuating the device, a maximum of 10 bars will be displayed. When the display
remains stable, the ultimate vacuum has been reached and the evacuation time begins.
The display of the ultimate vacuum (10 bars) depends on the vacuum pump's
performance and the atmospheric pressure.
Important note:
The DIGIMON manifold must be switched on before commissioning the attached
vacuum pump. The manifold must be switched on during evacuation.