Guidance and manufacturers declaration – electromagnetic
immunity
leva Pelvic Health System is intended for use in the
electromagnetic environment specified below. The
customer or the user of leva Pelvic Health System
should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
guidance
Conduc
ted RF
IEC
61000-4
-6
Radiate
d RF IEC
61000-4
-3
3 Vrms
150 kHz
to 80 MHz
3 V/m
80 MHz to
2,5 GHz
[V1] = 3 Vrms
[E1] =3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of leva Pelvic
Health System,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
where P is the
maximum output
power rating of the
transmitter in watts
(W) according to the
transmitter
manufacturer and d is
45