the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey (a), should be
less than the
compliance level in
each frequency
range. (b)
Interference may
occur in the
vicinity of
equipment
marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects, and people.
a. Field strengths from fixed transmitters, such as base stations
for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured
field strength in the location in which the leva Pelvic Health System
is used exceeds the applicable RF compliance level above, the leva
Pelvic Health System should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the leva Pelvic Health