Immunity test
IEC 60601 test level
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15kV air
±8kV contact
±15 kV air
±2kV for power
supply lines
±1 kV differential
mode
±2 kV for power
supply lines
30A/m 30A/m
± 1 kV line(s) to line(s)
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Electrical fast
transient/burst
IEC 61000-4-4
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
07
08
NOTE UT is the a.c. mains voltage prior to application of the test level.
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
Immunity test
Electromagnetic environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
Floors should be wood, concrete or ceramic tile.
If floor are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of XD100 Ventilation Disinfector requires contin-
ued operation during power mains interruptions,
it is recommended that XD100 Ventilation
Disinfector be powered from an uninterruptible
power supply or
a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.