Full face mask
English 1
Disinfection guide Clinical use only
This guide is intended for multi-patient re-use of the AirFit
F30i full face mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning
instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.
Mask component
1
Cleaning - Mild alkaline, anionic detergent
eg, Alconox
Validated number of cycles
Thermal Disinfection
167°F (75°C) for 30 minutes OR
194-199°F (90-93°C) for 1 to 10 minutes
Cushion
30
Frame / frame magnetic clips
30
Elbow body / elbow clip / elbow ring
30
Headgear
2
30
Soft wraps
3
1
May not be available in all regions. For full details regarding the correct use of this mask, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com.
2
Headgear includes headgear magnetic clips.
3
Not intended for multi-patient re-use.
3
D
C
A
B
1
2
4
A Elbow body
1 Elbow clip
2 Elbow ring
B Frame
3 Frame connector
4 Frame magnetic clip
C Cushion
D Headgear
Additional disassembly
a b c

Other manuals for ResMed AirFit F30i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirFit F30i and is the answer not in the manual?

ResMed AirFit F30i Specifications

General IconGeneral
TypeFull Face Mask
Cushion MaterialSilicone
Size OptionsSmall, Medium
HeadgearIncluded
Vent TypeQuietAir vent
Compatible MachinesResMed AirSense 10, AirCurve 10, AirMini
Tubingstandard 22mm tubing required
Recommended CleaningDaily cleaning with mild soap and water
Headgear MaterialNylon/Spandex

Related product manuals