Disinfection guide Clinical use only
This guide is intended for multi-patient re-use of the AirFit
™
F30i full face mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning
instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.
Mask component
1
Cleaning - Mild alkaline, anionic detergent
eg, Alconox
Validated number of cycles
Thermal Disinfection
167°F (75°C) for 30 minutes OR
194-199°F (90-93°C) for 1 to 10 minutes
• Cushion
30
• Frame / frame magnetic clips
30
• Elbow body / elbow clip / elbow ring
30
• Headgear
2
30
• Soft wraps
3
– –
1
May not be available in all regions. For full details regarding the correct use of this mask, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com.
2
Headgear includes headgear magnetic clips.
3
Not intended for multi-patient re-use.
A Elbow body
1 Elbow clip
2 Elbow ring
B Frame
3 Frame connector
4 Frame magnetic clip
C Cushion
D Headgear
Additional disassembly
a b c