40
English 1
Disinfection guide Clinical use only
This guide is intended for multi-patient re-use of only the Sleep Lab Mask (SLM) variant of the AirFit F40 mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer
to the User Guide for cleaning instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures. Potable water with hardness
≤150 mg/L is recommended for all cleaning, rinsing and disinfection steps.
Mask component
1
Cleaning
Validated number of cycles
Alconox™ (Mild alkaline, anionic detergent)
Thermal Disinfection
Manual
167°F (75°C) for 30 minutes
194-199°F (90-93°C) for 1 to 10 minutes
Cushion
30
Frame
Multi-hole vent ring
Elbow and short tube
Headgear
QuietAir™ vent ring -
Not intended for multi-patient use
1
May not be available in all regions. For full details regarding the correct use of these masks, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com
A Elbow and short tube
B Frame
C Cushion
D Headgear
1 Anti-asphyxia valve (AAV)
2 Side button
3 Frame stud hole
4 Frame magnet
5 QuietAir vent ring
6 Multi-hole vent ring
7 Cushion stud
8 Magnetic clip
9 Lower headgear strap
10 Upper headgear strap
11 Top headgear strap