Respironics Solo - Specifications; Device Dimensions and Weight; Operating Environment; Power Requirements

Respironics Solo
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4
Specifications
Dimensions: 10.0" x 5.55" x 4.25" (25x14x11 cm)
Weight: < 5 lb. (2.27 kg)
Temperature: 41 - 104° Fahrenheit (5 - 40° Celsius)
Humidity: 30-75% Non-condensing
Atmospheric Pressure: 83 - 102 kPascals (storage: 50 - 102 kP)
The electrical installation of the relevant room must comply with
IEC requirements.
115-120 VAC / 230-240 VAC, 50/60 Hz, 0.5 A max.
12 VDC, 2 A DC max.
Class II, Type BF
Degree of protection against harmful ingress of water: ordinary
(IPXO)
Mode of Operation: Continuous
Fuses: Line Voltage Fuse Type
115 VAC (115-120 VAC) T 315 mA, 5 x 20 mm
230 VAC (230-240 VAC) T 200 mA, 5 x 20 mm
Use only Respironics fuses supplied by your home care provider.
Device Size
Product Use
and Storage
Environment
Power
Requirements
IMPORTANT!
DC Power
Type B
Fuse
Class II
(Double Insulated)
Attention, consult
accompanying documents
!
SYMBOL KEY
®
®
1001 Murry Ridge Drive
Murrysville, Pennsylvania
15668-8550 U.S.A.
®
11, Place Royale
44000 Nantes France