RT300 Supine Service Guide
© Restorative Therapies Inc. 2017. Issue 2.1
RT300
Supine
Service
Guide
for build 2
Caution for RT300 users in the USA:
Federal law restricts this device to sale by, or on the order of, a
practitioner licensed by the law of the state in which he or she
practices to use or order the use of the device.

Other manuals for Restorative Therapies RT300

Question and Answer IconNeed help?

Do you have a question about the Restorative Therapies RT300 and is the answer not in the manual?

Restorative Therapies RT300 Specifications

General IconGeneral
BrandRestorative Therapies
ModelRT300
CategoryMedical Equipment
LanguageEnglish

Summary

Preventative Maintenance Procedures

Procedures After Each Use

Details the cleaning steps for patient contact areas and components after each use of the RT300 Supine device.

Procedures After Each Day

Outlines daily checks, including confirming crank tightness and internet connection status for the RT300 Supine.

Procedures After Each Week

Covers weekly maintenance tasks such as confirming SAGE mounting tightness and inspecting foot/leg straps.

Procedures After 50 Hours of Use

Details maintenance required every 50 hours, including inspecting cords and checking system hours.

Procedures Annually

Describes annual maintenance tasks like general inspection and testing of safety switches.

Ergometer Calibration

Provides a step-by-step guide for recalibrating the RT300 Supine ergometer annually.

RT300 Supine Diagnostics

Power Up Issues

Troubleshooting steps for RT300 Supine not powering up, checking AC power and fuses.

Error: Stimulator Disconnected

Guidance for when the RT60 switch is off or the power cable is not fully inserted.

Error: Ergometer Disconnected

Troubleshooting for 19V supply failure or tablet running on battery.

Error: Crank Rotation Not Detected

Addresses issues with SAGE tablet pairing, 24V supply, or motor/wiring faults.

Drive Unit Belt Issues

Troubleshooting for drive unit belt problems like breakage, looseness, or improper tension.

SAGE Software Issues

Steps for resolving SAGE not powering up or software not starting issues.

Stimulation: Cannot Find RT60

Troubleshooting when SAGE cannot find the RT60, checking power and pairing.