305978H02 Rev A
© 2003
All rights reserved
Printed in U.S.A.
Ultima II and Elite II
Laboratory CO
2
Incubators
Installation and Operation Manual
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Operating Standards . . . . . . . . . . . . . . . . . . . . . . . . 2
Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Initial Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Key Switch and Control Operation . . . . . . . . . . . . . 8
Control Systems Theory . . . . . . . . . . . . . . . . . . . . 13
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . 19

Questions and Answers

  • Y
    ynavarroSep 13, 2025
    What could be the reasons for Digital CO2 display and Fyrite read 0% CO2, Setpoint is okay, CO2 alarm and light are activated, CO2 inject light is on in REVCO Ultima II Accessories?
    • N
      Natasha ThomasSep 13, 2025
      If the digital CO2 display and Fyrite read 0% CO2, the setpoint is okay, the CO2 alarm and light are activated, and the CO2 inject light is on, it may be due to loss of CO2 supply, a defective CO2 solenoid, or a blocked CO2 inline filter.
  • A
    Amanda AlvarezAug 25, 2025
    What to do if the Add-Water indicator is illuminated and alarm is sounding on REVCO Accessories?
    • J
      jacqueline62Aug 25, 2025
      If the 'Add-Water' indicator is illuminated and the alarm is sounding, take the following steps: * Avoid filling the incubator with de-ionized water. * Inspect the floor around the unit for any signs of a water leak. * Ensure the incubator is placed on a level surface. * Verify that water is visible at the weep hole.
  • K
    Kevin MontgomeryAug 26, 2025
    Why is water leaking from the REVCO Ultima II cabinet?
    • K
      Kelly GarciaAug 26, 2025
      If water is leaking from the cabinet, it might be because the water jacket is overfilled. Visually check the water level and drain some water from the water jacket if needed.
  • M
    marcus88Sep 23, 2025
    What could be the reasons for CO2 setpoint changes by itself in REVCO Accessories?
    • J
      Jennifer WilliamsSep 23, 2025
      If the CO2 setpoint changes by itself, it may be due to a defective setpoint potentiometer, a defective CO2 control, or EMI or RFI.
  • H
    hlongSep 14, 2025
    What could be the reasons for Digital CO2 display locks up and shows some random number or a decimal point in REVCO Ultima II?
    • J
      James WalkerSep 14, 2025
      If the digital CO2 display locks up and shows some random number or a decimal point, it may be due to a defective display board, a defective output to the display board, or faulty interconnect wiring.
  • P
    Penny SnowSep 23, 2025
    What could be the reasons for Actual CO2 is higher than the setpoint and display, Re-zeroing helps, but the symptom returns in REVCO Accessories?
    • M
      Megan RobinsonSep 23, 2025
      If the actual CO2 is higher than the setpoint and display, re-zeroing helps, but the symptom returns, it may be due to a defective CO2 sensor or incorrect calibration of the CO2 control.
  • C
    Candace CampbellSep 6, 2025
    What to do if the temperature is over-shooting during the day in REVCO Ultima II?
    • R
      Robert ChurchSep 6, 2025
      If the temperature is over-shooting during the day, avoid opening the door more frequently than at five minute intervals and check the incubator location.
  • O
    ohortonSep 7, 2025
    What to do if the CO2 concentration measurement does not equal the display on REVCO Accessories?
    • V
      Victoria RodriguezSep 7, 2025
      If the CO2 concentration measurement does not equal the display, check the setpoint and verify Fyrite analyzer operation.
  • M
    Melissa GilbertSep 23, 2025
    What could be the reasons for CO2 alarm sounds while CO2 level is controlling at setpoint in REVCO Ultima II Accessories?
    • R
      Riley SmithSep 23, 2025
      If the CO2 alarm sounds while the CO2 level is controlling at the setpoint, it may be due to excessive EFI or EMI near the cabinet, a faulty grounding circuit, or a defective CO2 control.
  • C
    Christopher AlvarezAug 31, 2025
    Why is the CO2 display drifting on REVCO Accessories?
    • T
      thompsonsamanthaAug 31, 2025
      If the CO2 display is drifting, close the glass door for at least ten minutes and check the CO2 gas supply.

REVCO Ultima II Specifications

General IconGeneral
BrandREVCO
ModelUltima II
CategoryAccessories
LanguageEnglish