EasyManua.ls Logo

Reveal AURA - 2.5 Warnings Related to RF Energy Emission

Default Icon
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator Manual
This equipment is for use only by qualied medical professionals trained in the particular
technique to be performed.
Connect the power cord to a properly polarized and grounded power source with the
frequency and voltage characteristics that match those listed on the back of the unit.
Connect the System power cord to a properly grounded receptacle.
Do not use power plug adapters.
Always turn o and unplug the device before cleaning.
The patient should not come into contact with metal parts which are earthed or which have
an appreciable capacitance to earth. The use of anti-static sheeting is recommended for
this purpose.
Treatment bed or chair should not be electric. Use the lowest output setting necessary to
achieve the desired treatment eect.
The higher RF is applied, the greater the possibility of unintended thermal damage. Failure
of the equipment could result in an unintended increase of output power.
To avoid the RISK of electric shock, this equipment must only be connected to a SUPPLY
MAINS with protective earth. Do not use the AURA™ on patients with pacemakers or
internal debrillators.
2.5 Warnings Related to RF Energy Emission
As required by national and international
regulatory agencies, appropriate warning and
information labels have been attached in specic
locations on the instrument as identied below.
Figure 2-1: AURA System Rear Panel with Label.
2.6 System and Handpieces Labels
7