Air Quality: the system should operate in a non-corrosive atmosphere. Corrosive materials
such as acids can damage electrical wiring, electronic components and the surfaces of
optical components. Air-borne dust particles should be kept to a minimum. Dust particles
absorb light/laser energy and heat up. Hot particles located on the optical lenses can
damage them. Metallic dust is destructive to electrical equipment.
Temperature: to ensure that the system performs optimally, maintain room temperature
between 20°C and 25°C (68°F - 77°F) and relative humidity of less than 80%. Note: When
the system is used intensively, it emits heat. Therefore, it is recommended to install air
conditioning in the room in which the system will be used.
3.3 Environmental Requirements
Space should be allocated with
adequate ventilation and free airow.
The working area for the system should
be prepared according to the system
dimensions presented in the Figure. In
order to guarantee proper ventilation,
always keep the sides of the system at
least 20” (0.5m) from the wall or from
other obstructions to air ow. After
positioning the system, lock the breaks
on the front wheels by pressing the
pedals on top of each wheel.
3.4 Space and Positioning
1 m
0.44 m
0.33 m