Operation Manual RHE 07, 08, 11
Page
25
d:
The d-values are given in promille (‰) of the time period of the base frequency.
Left is shown the phase shift of the last stored zeroing and right is the actual measured phase
shift.
Press button Vary to see only the phase shift changes..
In the case of zero flow, this value Vary should be very low.
ADChannel1:
The Analog input value is displayed after conversion by the 10 bit analog/digital converter. Value
range is 0 - 1023 steps . This channel is in use for sensor temperature measurement.
ADChannel2:
The second analog input channel has the same technical specifications as
ADChannel1. It is a channel for special use (see refer. 2.1.3.5.2 analog input)
RTime:
Electronic run time counter in days and hours.
Mass:
Second, non resetable totalizer.
2.1.4.3. Sensor basic level programming (BASIS SETUP)
To enter this menu you have to input the password.
The password is:
- Press 3 times key #1
- Press 2 times key #2
- Press 1 time key #3
After pressing a key an asterisk
* is being displayed.
Inside this menu you have to enter sensor specific data like meter size, type of connection (serial or
parallel) and maximum sensor operating temperature.