Err 3 Temperature The temperature detected by the PT 100 inside the sensor RHM is
outside of the allowed range (-154 .. 360 °C) or temperature is
outside of the allowable operating temperature, adjusted in
MaxTemp menu.
Possible causes:
PT 100 defective or circuit open or shortened
defective component on safety or amplifier board
defective analog to digital converter inside
defective voltage supply (component U7 or U9)
temperature measurement not properly adjusted
Err 4 Parameter Error on parameter check. Error occurred during parameter transfer from
EEProm to RAM memory. Calculated checksum is different from backuped
checksum.
Err 5 RAM Error on RAM check.
Err 6 ROM Error on ROM check. Defective EEProm storage cell detected
calculated checksum is different from programmed checksum.
Err 7 EEProm No EEProm reading or writing possible.
Err 8 Division Calibration error. Internal calculation overflow.
verify calibration parameter setting.
Err 9 Stack Stack memory too small. Reduce number of measurement
periods (gates) in calibration parameter setup (IntGates).
Err 10 A/DChan2 Defective analog input, or input voltage outside range
(0 - 5 Volt). Check input voltage
analog to digital converter defective.