What does LoW mean on Rice Lake SCT-2200 Advanced Transmitter?
- JJessica AlvarezSep 9, 2025
The 'LoW' error indicates that the net weight is less than the minimum necessary for printing or totalization.
What does LoW mean on Rice Lake SCT-2200 Advanced Transmitter?
The 'LoW' error indicates that the net weight is less than the minimum necessary for printing or totalization.
Why can't I print when the weight is in underload or overload on my Rice Lake Accessories?
Printing issues may arise if the weight is either in underload or overload. This can also occur if the weight exceeds the maximum capacity by 9 divisions or falls 100 divisions below the gross zero.
What to do if Rice Lake SCT-2200 Advanced shows Er-37 error?
The Er-37 error indicates that the scale must be calibrated. You may need to perform a technical default parameter (dEFAu) before proceeding.
What causes Er-11 error on Rice Lake SCT-2200 Advanced Transmitter?
The Er-11 error indicates a calibration error where the sample weight used was too small. To resolve this, use a weight equal to at least half of the scale's capacity.
Why does my Rice Lake SCT-2200 Advanced display bUSy?
The 'bUSy' message indicates that the PRN serial port is occupied, or the instrument is waiting to transmit a print job to a PC.
What does UnStAb mean on Rice Lake Transmitter?
The 'UnStAb' message means that you are trying to print with an unstable weight.
What does no in mean on Rice Lake SCT-2200 Advanced Transmitter?
The 'no in' message indicates a second attempt to acquire the input weight (input/output mode, set as in.out).
What does no.0.UnS mean on Rice Lake SCT-2200 Advanced?
The 'no.0.UnS' message indicates that the weight did not exceed net 0 or was not stable.
What does NEt.Err mean on Rice Lake Transmitter?
The 'NEt.Err' error means you are trying to print with a negative net weight (equal or less than 0).
What does PrEC mean on Rice Lake SCT-2200 Advanced Transmitter?
The 'PrEC' message appears when trying to calibrate a point without first having confirmed the number of calibration points.
Brand | Rice Lake |
---|---|
Model | SCT-2200 Advanced Series |
Category | Transmitter |
Language | English |
General safety precautions to prevent injury and equipment damage.
Allows selecting various operational functions for the instrument.
Provides a hierarchical overview of the instrument's setup menus and parameters.
Configures the instrument's functional modes and associated parameters.
Step-by-step guide for calibrating the scale.
Configurations for primary, secondary, and repeater roles in multi-instrument systems.
Lists and explains instrument error messages for problem diagnosis.