Richard Wolf E2085.00 Series - User Manual

Richard Wolf E2085.00 Series
6 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
EE
(LOF)
MODEL
E2085.00
SERIES
BIPOLAR
ELECTROSURGICAL
GENERATOR
OPERATION
MANUAL
This
instruction
manual
covers
the
functions
of
units
E2085.00B
(black
faceplate)
and
E2085.00
(cream
faceplate).
Question and Answer IconNeed help?

Do you have a question about the Richard Wolf E2085.00 Series and is the answer not in the manual?

Overview

The Richard Wolf Model E2085.00 is a mini-bipolar electrosurgical generator designed for surgical procedures requiring bipolar current. This device is available in two versions: E2085.00B (black faceplate) and E2085.00 (cream faceplate).

Function Description:

The E2085.00 provides controlled bipolar electrosurgical current for various surgical applications. A key feature is the integrated ammeter, which offers the surgeon a visual indication of the operation's progress. During procedures like tubal ligation, the ammeter typically shows a decrease in current flow as the ligation progresses. The unit also emits an audible tone and illuminates a Bipolar Indicator Light to confirm active functioning.

Important Technical Specifications:

  • Physical Dimensions:
    • Height: 5.0 inches
    • Width: 12.5 inches
    • Depth: 12.0 inches
    • Weight: 16 lbs.
  • Electrical Specifications:
    • Power Supply: 120 V AC, 60 Hz
    • Inside Fuse: 2.5 A/250 V SLO-BLO
    • Outside Fuse: 3 A/250 V
    • Cord Plug: Hospital Grade Hubbell
    • Line Cord: SJT 18/3-8ft. long
    • Bipolar Output: 50 Watts
    • Maximum Power: 200 Watts

Usage Features:

The E2085.00 is designed for straightforward operation with several user-friendly features:

  • Controls and Indicators:
    • Power Switch (1): Turns the unit on/off.
    • Footswitch Receptacle (2): Connects to the footswitch for activation.
    • Bipolar Receptacle (3): Connects to the active bipolar cable.
    • Bipolar Current Control (4): Adjusts the current output.
    • Ammeter (5): Provides visual feedback on current flow.
    • Bipolar Indicator Light (6): Illuminates when the unit is functioning.
    • Bumper Trim (7): A protective feature.
    • Line Cord (8): For connecting to the AC power supply.
    • Fuse (9): 3 A/250 V fuse for protection.
    • Handle (10): For easy transport.
  • Operating Instructions (Bipolar):
    1. Ensure all cables are in good condition and free of moisture.
    2. Connect the footswitch cable to the Footswitch Receptacle (2) and position it for the surgeon.
    3. Insert the active bipolar cable into the Bipolar Receptacle (3). Do not connect the active cable to instruments until the unit is ready.
    4. Depress the Power Switch (1).
    5. Depress the footswitch when power is needed.
    6. Monitor the Ammeter (5) for visual indication of operation progress.
    7. The unit's tone and the Bipolar Indicator Light (6) confirm active functioning.
  • Safety Precautions:
    • Always unpack carefully and inspect for damage.
    • Test the unit prior to surgical procedures.
    • NEVER operate in the presence of flammable anesthetics.
    • NEVER remove the cover due to electric shock hazard; refer servicing to qualified personnel.
    • Always replace blown fuses with the same type and rating.
    • Ensure proper electrical grounding.
    • Do not block ventilation vents.
    • WARNING: Do not modify the bipolar cable in any way, as modifications void all warranties and are done at the user's own risk.

Maintenance Features:

  • Fuse Replacement: To replace the fuse, push the Fuseholder Cap (9) in and to the left, replace the blown fuse with a 3 A/250 V fuse, then push the cap in and turn to the right.
  • Troubleshooting Chart:
    • No Bipolar Output, No Indicator Light or Audible Tone:
      • Verify connection to AC power supply.
      • Inspect line cord for damage.
      • Check for blown 3A fuse.
      • Ensure footswitch plug is properly inserted and secured.
    • No Bipolar Output, Indicator Lamp Lights and Tone Sounds but No Power Output:
      • Check cables, plugs, and insulation for good condition.
      • Ensure cable between bipolar forceps and receptacle is properly inserted and secured.
      • Check instrument tips for encrusted tissue.
      • Verify an adequate power setting is selected.
  • Warranty and Customer Service: Richard Wolf unconditionally guarantees instruments to be free from defects in materials and workmanship under normal use and service. Repairs are handled as customer accommodation. Any repairs, maintenance, or servicing by unauthorized personnel will void the warranty.
  • Repair Policy: Defective merchandise will be repaired or replaced at no charge if delivered prepaid. When returning instruments for repair, it is recommended to use the original packaging and insure the instrument.
  • IMPORTANT: For general safety and health reasons, all instruments must be cleaned and sterilized before returning for repair. Instruments received in an unsanitary condition will be returned immediately to their point of origin, prepaid.
  • Customer Service Contact:
    • In Illinois: (708) 913-1113
    • Outside of Illinois: (800) 323-WOLF
    • Repair Service: 1 (800) 323-2160