How do I use Multi-Flo in Remote
Control mode on Med-eBase?
To use the remote control mode the user will need to install the Gadget
Serial V2.4 Driver. The Gadget serial driver is available at rigelmedical.com in
the download section. To install the gadget serial driver the Multi-Flo needs
to be turned on and connected to the PC using the USB serial cable.
Equipment required:
- Multi-Flo
- PC running Med-eBase V2.4 or higher
- Gadget serial driver install files
- USB Serial cable or memory stick
1) On the Multi-Flo menu scroll down and select remote control mode. The
Multi-Flo will then try to connect with the PC.
2) The PC will attempt to download the install drivers for the gadget serial.
As the driver is located on the PC and cannot be found on the internet this
step is not required. Click on the icon when it appears as seen below.
3) Select “skip obtaining driver software” which is blue and underlined to
stop widows looking for driver as the Rigel device won’t be able to connect
to the PC while the computer is performing this task. Select yes on the
driver software installation screen to skip the driver update.
Question and Answer IconNeed help?

Do you have a question about the Rigel MULTI-FLO and is the answer not in the manual?

Summary

Using Multi-Flo in Remote Control Mode on Med-eBase

Initial Setup and Connection

Install the Gadget Serial V2.4 Driver and connect the Multi-Flo to the PC via USB serial cable.

Start Remote Control Mode

On the Multi-Flo, scroll to and select remote control mode to initiate PC connection.

Skip Driver Update

Select 'skip obtaining driver software' and confirm to prevent driver update interference.

Configure Gadget Serial Driver

Access Control Panel, update driver by browsing for files in the correct folder.

Verify Installation and Download

Confirm driver installation and download data from the Multi-Flo via Med-eBase.

Summary

Using Multi-Flo in Remote Control Mode on Med-eBase

Initial Setup and Connection

Install the Gadget Serial V2.4 Driver and connect the Multi-Flo to the PC via USB serial cable.

Start Remote Control Mode

On the Multi-Flo, scroll to and select remote control mode to initiate PC connection.

Skip Driver Update

Select 'skip obtaining driver software' and confirm to prevent driver update interference.

Configure Gadget Serial Driver

Access Control Panel, update driver by browsing for files in the correct folder.

Verify Installation and Download

Confirm driver installation and download data from the Multi-Flo via Med-eBase.

Overview

The Rigel Medical Multi-Flo is a versatile medical device designed for testing and calibration, particularly in conjunction with the Med-eBase software. Its primary function is to facilitate remote control operation, allowing users to manage and interact with the device through a connected computer. This capability streamlines testing procedures and data management, enhancing efficiency in medical equipment maintenance and calibration environments.

Function Description

The Multi-Flo, when operated in Remote Control mode, integrates seamlessly with Med-eBase V2.4 or higher. This integration enables comprehensive control over the Multi-Flo's functions directly from a PC, eliminating the need for constant manual interaction with the device's interface during testing. The core of this functionality relies on the installation of the Gadget Serial V2.4 Driver, which establishes the necessary communication link between the Multi-Flo and the computer. Once connected and configured, the Multi-Flo can be instructed to perform various tests and operations, with results and status updates displayed within the Med-eBase software. This remote control capability is particularly beneficial for automated testing sequences, data logging, and generating detailed reports, which are crucial for compliance and record-keeping in medical facilities. The device is designed to be robust and reliable, ensuring accurate and consistent performance during critical calibration and testing tasks. Its ability to connect via a standard USB serial cable makes it accessible for a wide range of users and computer setups.

Usage Features

Operating the Multi-Flo in Remote Control mode involves a straightforward setup process. Users begin by powering on the Multi-Flo and connecting it to a PC using the provided USB serial cable. The device's menu allows for easy selection of the "remote control mode," initiating the connection attempt with the computer. A key aspect of its usability is the driver installation process. While the PC may initially attempt to search for drivers online, the instructions guide users to "skip obtaining driver software" and instead locate the pre-downloaded Gadget Serial driver files. This ensures a quick and efficient installation without relying on internet connectivity.

Once the driver is installed, the Multi-Flo appears as an unspecified device in the computer's 'Devices and Printers' section. Users can then update the driver by browsing to the folder containing the gadget serial files. This step is crucial for establishing proper communication. After the driver is successfully installed, the Med-eBase software can be opened. Within Med-eBase, users can select "Download from Tester" and specify "Multi-Flo" as the instrument type, then choose "Gadget serial" as the COM port. This final configuration step enables the full remote control functionality. The system is designed to be user-friendly, with clear steps for installation and setup, minimizing potential complications. The ability to download data directly from the tester into Med-eBase facilitates efficient data management and analysis, allowing for the creation of comprehensive test reports. The Multi-Flo's interface, even when not in remote mode, is intuitive, allowing users to navigate through menus to select various operational modes. The device's design prioritizes ease of use, ensuring that technicians can quickly integrate it into their workflow.

Maintenance Features

While the provided information primarily focuses on the operational aspects of the Multi-Flo in remote control mode, several implicit maintenance considerations can be inferred. The reliance on driver installation and proper software configuration highlights the importance of keeping the Med-eBase software and associated drivers up-to-date. Regular checks for software updates from Rigel Medical or Seaward, the parent company, would ensure optimal performance and compatibility with evolving operating systems.

The use of a USB serial cable for connection implies that proper care of this cable is essential to maintain reliable communication. Damaged cables could lead to intermittent connections or complete loss of remote control functionality, necessitating replacement. Similarly, the physical integrity of the Multi-Flo device itself, including its ports and internal components, is crucial. Routine visual inspections for any signs of wear, damage, or loose connections would contribute to its longevity and consistent operation.

The instruction to close and reopen Med-eBase or rescan COM ports if the gadget serial is not initially located after driver installation suggests a troubleshooting step that users can perform. This indicates a design that anticipates and provides solutions for common connectivity issues, empowering users to resolve minor problems independently. For more complex issues, the availability of a support contact via the Seaward website underscores a commitment to ongoing product support and maintenance. This ensures that users have access to expert assistance when needed, minimizing downtime and maximizing the operational lifespan of the Multi-Flo. The overall design emphasizes a robust and maintainable system, where software and hardware work in concert to provide a reliable testing solution.

Rigel MULTI-FLO Specifications

General IconGeneral
Dimensions210 x 120 x 80 mm (8.3 x 4.7 x 3.1 in)
Weight1.2 kg (2.6 lbs)
TypeInfusion Pump
ConnectivityUSB
DisplayLCD
Power SupplyRechargeable battery or AC adapter
ChannelsSingle Channel
Operating Temperature15 to 35°C (59 to 95°F)
Storage Temperature-20 to 50°C (-4 to 122°F)