Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA3000 User Guide 2-37
VSWR
Enables or disables the VSWR measurement function. When the VSWR measurement
function is enabled, the user interface is divided into two sections automatically (the
lower section provides the measurement wizard). At this time, you can perform the
VSWR measurement according to the wizard. Press Meas Setup to set the
corresponding parameters.
Tip:
The VSWR bridge and tracking generator are required in VSWR measurement.
Therefore, this function is only available for RSA3045-TG/RSA3030-TG installed
with VSWR bridge. When the VSWR measurement is enabled, the tracking
generator is enabled automatically, and the backlight of TG on the front panel
turns on.